Flow visualization of arteriovenous anastomosis model using particle image velocimetry

被引:0
作者
Sasaki K. [1 ]
Shinne Y. [1 ]
Oku T. [2 ]
Yamauchi S. [2 ]
Motohashi Y. [2 ]
Sato T. [1 ,2 ]
Agishi T. [3 ]
机构
[1] Graduate School of Engineering, Toin University of Yokohama
[2] Department of Clinical Engineering, Faculty of Biomedical Engineering, Toin University of Yokohama
来源
Transactions of Japanese Society for Medical and Biological Engineering | 2020年 / 58卷
关键词
Arteriovenous fistula(AVF)model; Flow visualization; Particle Image Velocimetry(PIV); Shunt murmur; VascularAccess (VA);
D O I
10.11239/JSMBE.ANNUAL58.101
中图分类号
学科分类号
摘要
Shunt murmurs are believed to be generated when turbulent flow, which is generated as a large amount of blood passes quickly through the anastomosis site of the VA, causes the vascular wall to vibrate. In this report, the pseudo shunt murmurs generated in the AVF models with different diameter stenosis rates was measured, and the flow in these models were visualized using particle image velocimetry (PIV). As a result, in the several model with a low diameter stenosis rate, the flows were laminar, and the frequency component of the pseudo shunt murmurs were mainly composed of low frequency components. As a result, it was considered that the change in the flow in the blood vessels by stenosis may have an effect on the acoustic characteristics of the shunt murmurs. © 2020, Japan Soc. of Med. Electronics and Biol. Engineering. All rights reserved.