Multisensor Magnetic Scanning Microscope for Remanent Magnetic Field Measurements

被引:1
作者
Chaves, Joao F. [1 ]
Noris, Leosdan F. [1 ]
Yokoyama, Elder [2 ]
Osorio G., Fredy G. [1 ]
Mendoza, Leonardo A. F. [3 ]
Araujo, Jefferson F. D. F. [1 ]
机构
[1] Pontifical Catholic Univ Rio De Janeiro, Dept Phys, BR-22451900 Rio De Janeiro, Brazil
[2] Univ Brasilia, Dept Geosci, BR-70910900 Brasilia, Brazil
[3] State Univ Rio De Janeiro UERJ, Dept Elect Engn, BR-20550900 Rio De Janeiro, Brazil
关键词
Magnetic Scanning Microscopy; Delta-Mode; remanent fields; Hall sensor; Magnetoresistance sensor; PROBE; SAMPLE;
D O I
10.3390/s24072294
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Magnetic Scanning Microscopy (MSM) emerged with the aim of allowing the visualization of magnetic fields of a sample or material through scanning and proved particularly useful for geology, biomedicine, characterization of magnetic materials, and in the steel industry. In this regard, the reading system of an MSM was modified using a mu-metal magnetic shielding structure to analyze remanent fields. The MSM was adapted to perform readings using two different types of sensors. The sensitive area of the sensors was evaluated, and the HQ-0811 (AKM-Asahi KaseiTM Microdevices) and STJ-010 (Micro MagneticsTM) sensors were chosen, with the HQ-0811 standardized on Printed Circuit Boards (PCBs) to facilitate handling and increase the system's robustness. In the shielded chamber, two piezoelectric ANC-150 stepper motors (Attocube Systems) were used, arranged planarly, to allow the movement of the analyzed samples under the mounted sensors. To acquire data from the sensors, the Precision Current Source Model 6220 and the Nanovoltmeter Model 2182A (both from Keithley) were used, along with Keithley's Delta-Mode integrated system. To analyze the system's effectiveness, three distinct samples were analyzed for calibration, and a MATLAB program was written to analyze the images and extract the material's magnetization. Additionally, a rock sample from the Parnaiba Basin was mapped to demonstrate the system's capabilities.
引用
收藏
页数:16
相关论文
共 42 条
[21]   A scanning Hall probe microscope for large area magnetic imaging down to cryogenic temperatures [J].
Gregory, JK ;
Bending, SJ ;
Sandhu, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (10) :3515-3519
[22]   A Magnetic Sensor with Amorphous Wire [J].
He, Dongfeng ;
Shiwa, Mitsuharu .
SENSORS, 2014, 14 (06) :10644-10649
[23]   Ferrite Scanning Microscope Based on Magnetic Tunnel Junction Sensor [J].
Jaramillo, Gerardo ;
Chan, Mei-Lin ;
Milewski, John O. ;
Field, Robert D. ;
Horsley, David A. .
IEEE TRANSACTIONS ON MAGNETICS, 2012, 48 (11) :3677-3680
[24]   TUNNELING BETWEEN FERROMAGNETIC-FILMS [J].
JULLIERE, M .
PHYSICS LETTERS A, 1975, 54 (03) :225-226
[25]   Scanning SQUID susceptometers with sub-micron spatial resolution [J].
Kirtley, John R. ;
Paulius, Lisa ;
Rosenberg, Aaron J. ;
Palmstrom, Johanna C. ;
Holland, Connor M. ;
Spanton, Eric M. ;
Schiessl, Daniel ;
Jermain, Colin L. ;
Gibbons, Jonathan ;
Fung, Y. -K. -K. ;
Huber, Martin E. ;
Ralph, Daniel C. ;
Ketchen, Mark B. ;
Gibson, Gerald W., Jr. ;
Moler, Kathryn A. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (09)
[26]   Scanning magnetic tunnel junction microscope for high-resolution imaging of remanent magnetization fields [J].
Lima, E. A. ;
Bruno, A. C. ;
Carvalho, H. R. ;
Weiss, B. P. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (10)
[27]  
Marcon P, 2012, PR ELECTROMAGN RES S, P420
[28]   MAGNETIC IMAGING BY FORCE MICROSCOPY WITH 1000-A RESOLUTION [J].
MARTIN, Y ;
WICKRAMASINGHE, HK .
APPLIED PHYSICS LETTERS, 1987, 50 (20) :1455-1457
[29]  
melexis, MELEXIS (Inspired/Engineering) What Is the Hall-Effect?
[30]  
MicroMagnetics, SJT-010 Magnetic Microsensor with Socket Receptacle