Study on Mass Calibration Segmentation Algorithm of Ion Trap Mass Spectrometer

被引:0
|
作者
Pei H.-M. [1 ,2 ]
Liu M.-Y. [2 ]
Xu H.-Y. [2 ]
Zhang D. [2 ]
Xie J. [2 ]
Qu Z.-Y. [2 ]
Huang Z.-J. [2 ]
Jiang C.-Y. [2 ]
Meng Y.-Y. [2 ]
Dai X.-H. [2 ]
Fang X. [2 ]
Wang Q.-H. [1 ]
Jiang Y. [2 ]
机构
[1] Shenyang University of Chemical Technology, Shenyang
[2] National Institute of Metrology, Beijing
关键词
automatic calibration algorithm; linear ion trap; mass calibration; mass shift; mass spectrometer;
D O I
10.7538/zpxb.2022.0178
中图分类号
学科分类号
摘要
In order to ensure the measurement accuracy of ion trap mass spectrometer, it is necessary to perform mass calibration on the instrument when there is a mass shift. Mass calibration can be achieved by improving the stability of the electrical components or upgrading the hardware with significant technical costs. Compared with hardware optimization, mass calibration based on software algorithms can better utilize the performance of existing instruments. Mass calibration can be achieved by manually inputting standard samples or calibration solution parameters. However, mass axis calibration in this way still requires setting specific segmentation points based on the actual situation and manually adjusting them to achieve the desired accuracy. Therefore, to improve the low efficiency with manual segmentation during mass calibration, an automatic segmented calibration algorithm was proposed to improve the calibration accuracy and facilitate rapid operation. The algorithm used point by point fitting to determine whether the segment interval met the requirements and also determined each segment interval. At the same time, the mass axis was automatically segmented to reduce the mass deviation of each checkpoint. This method was compared with traditional method, different deviation thresholds were explored, and multiple instruments were selected to verify this method. The results showed that the algorithm can reasonably segment the mass axis and establish a new linear relationship between electrical parameters and mlz under existing conditions, allowing the deviation of each checkpoint to be controlled within a lower preset range ( + 0. 1 u), thus improving the instrument's calibration efficiency and measurement accuracy, reducing the difficulty of personnel operation. The significance of this study is that the algorithm only adjusts and optimizes the signals generated by the hardware without changing the structure of the instrument itself, which is helpful for improving the overall performance and stability of the instrument. Moreover, the principle and conclusion of the algorithm are independent of the experimental platform, so it can be applied to a variety of other types of mass spectrometers. For the insufficient points, firstly, the upper limit of accuracy provided by the algorithm itself is still determined by the hardware accuracy of the instrument, so it is only possible to calibrate and optimize the mass axis within a certain accuracy range. Secondly, because the calibration process requires a certain number of reference points, which poses certain requirements for the selection of calibration solutions. © 2023 Chinese Society for Mass Spectrometry. All rights reserved.
引用
收藏
页码:676 / 684
页数:8
相关论文
共 12 条
  • [1] WANG Yan, LIU Haipei, SUN Chuanqiang, HAN Wennian, JIA Mingzheng, JIANG Xue-hui, Quantitative analysis algorithm of liquid chromatography-triple quadrupole mass spec-trometry, Journal of Tianjin University (Science and Technology), 53, 6, pp. 617-625, (2020)
  • [2] DOROSHENKO V M, COTTER R J, LUB-MAN D M., Linear mass calibration in the quadrupole ion-trap mass spectrometer, Rapid Communications in Mass Spectrometry, 8, 9, pp. 766-776, (1994)
  • [3] SCHWARTZ J C, REMES P M., Methods and apparatus for calibrating ion trap mass spectrometers
  • [4] CHEN Huanwen, LI Ming, FEI Qiang, JIN Qinhan, Mass calibration of miniaturized ion trap [J], Chinese Journal of Analytical Chemistry, 10, pp. 1505-1508, (2005)
  • [5] YASUAKI T, TAKAYUKI N, MINORU S, YUICHIRO H., Method for mass calibration
  • [6] WANG Weimin, JIN Liuyu, XU Fuxing, DING Chuanfan, Effect of amplitude-scanning ac on the performance of ion trap mass spectrometry, Journal of Chinese Mass Spectrometry Society, 42, 5, pp. 942-950, (2021)
  • [7] MOXOM J, REILLY P T A, WHITTEN W B, RAMSEY J M., Double resonance ejection in a micro ion trap mass spectrometer, Rapid Communications in Mass Spectrometry, 16, 8, pp. 755-760, (2002)
  • [8] DENG Bo, The discussion on some problems with construction and application of calibration curve, Chinese Journal of Inorganic Analytical Chemistry, 1, 3, pp. 1-7, (2011)
  • [9] LIU Lei, JIANG You, LIU Meiying, DAI Xinhua, FANG Xiang, CHEN Dazhou, QIU Chun-ling, HUANG Zejian, Research on automatic adjustment of mass resolution in quadrupole mass spectrometry, Journal of Chinese Mass Spectrometry Society, 41, 2, pp. 118-124, (2020)
  • [10] XIE Jie, YI Keke, GAO Wenjing, LIU Hao, LI Jialian, HUANG Zejian, LIU Meiying, YU Qingni, WANG Hanwen, JIANG You, DAI Xinhua, FANG Xiang, Determination of 250HD in serum based on quadrupole-linear ion tandem mass spectrometric technique[J], Journal of Chinese Mass Spectrometry Society, 44, 1, pp. 13-24, (2023)