Development of an integrated dual-modality 3D bioluminescence tomography and ultrasound imaging system for small animal tumor imaging

被引:2
|
作者
Su, Shih-Po [1 ]
Yang, Yuan-Zhen [1 ]
Chiang, Huihua Kenny [1 ,2 ]
机构
[1] Natl Yang Ming Chiao Tung Univ, Dept Biomed Engn, Taipei 11221, Taiwan
[2] Natl Yang Ming Chiao Tung Univ, Biomed Engn Res & Dev Ctr, Taipei, Taiwan
来源
OPTICS EXPRESS | 2024年 / 32卷 / 04期
关键词
GUIDED RADIATION-THERAPY; FLUORESCENCE MOLECULAR TOMOGRAPHY; X-RAY; RECONSTRUCTION; MOUSE;
D O I
10.1364/OE.507659
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Ultrasound (US) is a valuable tool for imaging soft tissue and visualizing tumor contours. Taking the benefits of US, we presented an integrated dual-modality imaging system in this paper that achieves three-dimensional (3D) bioluminescence tomography (BLT) with multi-view bioluminescence images and 3D US imaging. The purpose of this system is to perform non-invasive, long-term monitoring of tumor growth in 3D images. US images can enhance the accuracy of the 3D BLT reconstruction and the bioluminescence dose within an object. Furthermore, an integrated co-registered scanning geometry was used to capture the fused BLT and US images. We validated the system with an in vivo experiment involving tumor-bearing mice. The results demonstrated the feasibility of reconstructing 3D BLT images in the tumor region using 3D US images. We used the dice coefficient and locational error to evaluate the similarity between the reconstructed source region and the actual source region. The dice coefficient was 88.5%, and the locational error was 0.4 mm when comparing the BLT and 3D US images. The hybrid BLT/US system could provide significant benefits for reconstructing the source of tumor location and conducting quantitative analysis of tumor size. (c) 2024 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
引用
收藏
页码:5607 / 5620
页数:14
相关论文
共 50 条
  • [31] 3D ultrasound imaging in frequency domain with 1D array transducer
    Han, Zhihui
    Peng, Hu
    Lu, Putian
    ULTRASONICS, 2017, 76 : 28 - 34
  • [32] Optimal configurations of an electromagnetic tracking system for 3D ultrasound imaging of pediatric hips - A phantom study
    Pham, Thanh-Tu
    Le, Lawrence H.
    Andersen, John
    Lou, Edmond H.
    MEDICAL ENGINEERING & PHYSICS, 2024, 131
  • [33] A Dual-Use Imaging System for Pre-Clinical Small Animal Radiation Research
    Li, Meng
    He, Xingchi
    Eslami, Sohrab
    Kang-Hsin, Ken
    Zhang, Bin
    Wong, John
    Iordachita, Iulian
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 6904 - 6907
  • [34] Holographic tomography: hardware and software solutions for 3D quantitative biomedical imaging
    Kus, Arkadiusz
    Krauze, Wojciech
    Makowski, Piotr L.
    Kujawinska, Malgorzata
    ETRI JOURNAL, 2019, 41 (01) : 61 - 72
  • [35] Electron tomography combining ESEM and STEM: A new 3D imaging technique
    Jornsanoh, P.
    Thollet, G.
    Ferreira, J.
    Masenelli-Varlot, K.
    Gauthier, C.
    Bogner, A.
    ULTRAMICROSCOPY, 2011, 111 (08) : 1247 - 1254
  • [36] 3D parallel-detection microwave tomography for clinical breast imaging
    Epstein, N. R.
    Meaney, P. M.
    Paulsen, K. D.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (12):
  • [37] Singular value decomposition analysis of a photoacoustic imaging system and 3D imaging at 0.7 FPS
    Roumeliotis, Michael B.
    Stodilka, Robert Z.
    Anastasio, Mark. A.
    Ng, Eldon
    Carson, Jeffrey J. L.
    OPTICS EXPRESS, 2011, 19 (14): : 13405 - 13417
  • [38] Aperture synthesis with a monochromatic metasurface imaging system for 3D near-field imaging
    Boyarsky, Michael
    Sleasman, Timothy
    Pulido-Mancera, Laura
    Diebold, Aaron V.
    Imani, Mohammadreza F.
    Smith, David R.
    PASSIVE AND ACTIVE MILLIMETER-WAVE IMAGING XXI, 2018, 10634
  • [39] Employing Line-Laser Illumination for 3D Fluorescence Imaging and Tomography in a Tri-Modal SPECT-CT-OT Imaging System
    Cao, Liji
    Peter, Joerg
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2178 - 2184
  • [40] Performance evaluation of a rotatory dual-head PET system with 90° increments for small animal imaging
    Meng, F.
    Zhu, S.
    Li, L.
    Wang, J.
    Cao, X.
    Cao, X.
    Chen, X.
    Liang, J.
    JOURNAL OF INSTRUMENTATION, 2017, 12