Unraveling the influence of CRISPR/Cas13a reaction components on enhancing trans-cleavage activity for ultrasensitive on-chip RNA detection

被引:2
|
作者
He, Qian [1 ,2 ]
Chen, Qun [2 ]
Lian, Lijin [2 ]
Qu, Jiuxin [3 ]
Yuan, Xi [2 ]
Wang, Chuhui [2 ]
Xu, Lidan [2 ,4 ]
Wei, Jiazhang [5 ]
Zeng, Shaoling [6 ]
Yu, Dongmei [7 ]
Dong, Yuhan [2 ]
Zhang, Yongbing [8 ]
Deng, Lin [9 ]
Du, Ke [10 ]
Zhang, Canyang [2 ]
Pandey, Vijay [2 ]
Gul, Ijaz [1 ,2 ]
Qin, Peiwu [1 ,2 ]
机构
[1] Hangzhou Dianzi Univ, Sch Commun Engn, Hangzhou 310018, Zhejiang, Peoples R China
[2] Tsinghua Univ, Tsinghua Shenzhen Int Grad Sch, Inst Biopharmaceut & Hlth Engn, Shenzhen 518055, Guangdong, Peoples R China
[3] Shenzhen Third Peoples Hosp, Clin Lab, Shenzhen 518115, Guangdong, Peoples R China
[4] Harbin Med Univ, Lab Med Genet, Harbin 150081, Peoples R China
[5] Guangxi Acad Med Sci, Peoples Hosp Guangxi Zhuang Autonomous Reg, Dept Otolaryngol & Head & Neck, 6 Taoyuan Rd, Nanning 530021, Peoples R China
[6] Shenzhen Customs, Anim & Plant Inspect & Quarantine Technol Ctr, Shenzhen, Peoples R China
[7] Shandong Univ, Sch Mech Elect & Informat Engn, Weihai 264209, Shandong, Peoples R China
[8] Harbin Inst Technol Shenzhen, Shenzhen 518055, Peoples R China
[9] Shenzhen Bay Lab, Inst Mol Physiol, Shenzhen 518132, Peoples R China
[10] Univ Calif Riverside, Chem & Environm Engn, Riverside, CA USA
基金
中国国家自然科学基金;
关键词
CRISPR/Cas13a; TIRFM; RNA detection; Amplification-free; crRNA combination; NUCLEIC-ACID DETECTION; SARS-COV-2;
D O I
10.1007/s00604-024-06545-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The CRISPR/Cas13 nucleases have been widely documented for nucleic acid detection. Understanding the intricacies of CRISPR/Cas13's reaction components is pivotal for harnessing its full potential for biosensing applications. Herein, we report on the influence of CRISPR/Cas13a reaction components on its trans-cleavage activity and the development of an on-chip total internal reflection fluorescence microscopy (TIRFM)-powered RNA sensing system. We used SARS-CoV-2 synthetic RNA and pseudovirus as a model system. Our results show that optimizing Mg2+ concentration, reporter length, and crRNA combination significantly improves the detection sensitivity. Under optimized conditions, we detected 100 fM unamplified SARS-CoV-2 synthetic RNA using a microtiter plate reader. To further improve sensitivity and provide a new amplification-free RNA sensing toolbox, we developed a TIRFM-based amplification-free RNA sensing system. We were able to detect RNA down to 100 aM. Furthermore, the TIRM-based detection system developed in this study is 1000-fold more sensitive than the off-coverslip assay. The possible clinical applicability of the system was demonstrated by detecting SARS-CoV-2 pseudovirus RNA. Our proposed sensing system has the potential to detect any target RNA with slight modifications to the existing setup, providing a universal RNA detection platform.
引用
收藏
页数:12
相关论文
共 41 条
  • [41] Entropy-driven assisted T7 RNA polymerase amplification-activated CRISPR/Cas13a activity for SARS-CoV-2 detection in human pharyngeal swabs and environment by an electrochemiluminescence biosensor
    Wei, Jihua
    Song, Zichun
    Cui, Jiuying
    Gong, Yuanxun
    Tang, Qianli
    Zhang, Kai
    Song, Xinlei
    Liao, Xianjiu
    JOURNAL OF HAZARDOUS MATERIALS, 2023, 452