Design and performance testing for a vibration isolator of Raman retro-reflection mirror

被引:0
|
作者
Ru N. [1 ,2 ,3 ]
Li C. [2 ]
Wang Y. [3 ]
Zhang L. [3 ]
Fan S. [2 ]
机构
[1] Center for Advanced Measurement Science, National Institute of Metrology, Beijing
[2] School of Instrumentation Science and Opto-Electronics Engineering, Beihang University, Beijing
[3] Key Laboratory for Metrology, Changcheng Institute of Metrology and Measurement, Beijing
来源
Zhendong yu Chongji/Journal of Vibration and Shock | 2019年 / 38卷 / 19期
关键词
Atomic interference gravimeter; Raman retro-reflection mirror; Scanning laser interferometer; Simulation calculation; Vibration isolator;
D O I
10.13465/j.cnki.jvs.2019.19.030
中图分类号
学科分类号
摘要
Here, in order to improve Raman light phase stability of an atomic interference gravimeter and reduce effects of Raman retro-reflection mirror's vibration noise on measurement accuracy of the atomic interference gravimeter, a vibration isolator for small type atomic interference gravimeter's Raman retro-reflection mirror was designed and fabricated. Its vibration isolation performance was verified using COMSOL Multiphysics finite element analysis and laser interferometer tests. According to the requirements of atomic interference gravimeter for noise control, the vibration reduction effect of the vibration isolator within the range of 60-200 Hz was simulated. Taking the simulation results as the test guide, a scanning laser interferometer and a standard shaking table were used to measure vibration of a Raman retro-reflector loaded on the vibration isolator, the measurement results agreed well with the simulated ones. © 2019, Editorial Office of Journal of Vibration and Shock. All right reserved.
引用
收藏
页码:198 / 201
页数:3
相关论文
共 16 条
  • [1] Peters A., Chung Y.K., Chu S., Measurement of gravitational acceleration by dropping atoms, Nature, 400, (1999)
  • [2] McGuirk J.M., Snadden M.J., Kasevich M.A., Large area light-pulse atom gravimetry, Phys Rev Lett, 85, pp. 4498-4501, (2000)
  • [3] Snadden M., McGuirk J., Bouyer P., Et al., Measurement of the earth's gravity gradient with an atom gravimeter-based gravity gradiometer, Phys Rev Lett, 81, pp. 971-974, (1998)
  • [4] McGuirk J.M., Foster G.T., Fixler J.B., Et al., Sensitive absolute-gravity gradiometry using atom gravimetry, Phys Rev A, 65, (2002)
  • [5] Sorrentino F., Bodart Q., Cacciapuoti L., Et al., Sensitivity limits of a Raman atom gravimeter as a gravity gradiometer, Phys Rev A, 89, (2014)
  • [6] Biedermann G.W., Wu X., Deslauriers L., Et al., Testing gravity with cold-atom gravimeters, Phys Rev A, 91, (2015)
  • [7] Moler K., Weiss D.S., Kasevich M., Et al., Theoretical analysis of velocity-selective Raman transitions, Phys Rev A, 45, pp. 342-348, (1992)
  • [8] Ning R., Li Z., Yu W., Et al., The application of Raman laser in gravity measurement and metrology, Proc SPIE 9297, International Symposium on Optoelectronic Technology and Application 2014: Laser and Optical Measurement Technology
  • [9] and Fiber Optic Sensors, (2014)
  • [10] Alain M., Marion J., Matthias B., Et al., Vibration-induced phase noise in Mach-Zehnder atom interferometers, Applied Physics B-Lasers and Optics, 84, (2006)