Real-Time In-Situ Calibration for Angle Measuring Sensor Based on Autocollimator

被引:0
作者
Zhang W. [1 ,2 ]
Zhu H. [3 ]
Li M. [1 ,2 ]
Guo Z. [1 ,2 ]
Guo M. [1 ,2 ]
机构
[1] Institute for Interdisciplinary Quantum Information Technology, Jilin Engineering Normal University, Changchun, 130052, Jilin
[2] Jilin Engineering Laboratory for Quantum Information Technology, Changchun, 130052, Jilin
[3] School of Electronics and Information Engineering, Changchun University of Science and Technology, Changchun, 130022, Jilin
来源
Zhongguo Jiguang/Chinese Journal of Lasers | 2019年 / 46卷 / 04期
关键词
Angle measurement; Angle transducer; Error analysis; Measurement; Real-time calibration;
D O I
10.3788/CJL201946.0404005
中图分类号
学科分类号
摘要
An in-situ calibration method based on the autocollimator is proposed. Based on the principle of circumferential closure and the nature of Fourier series, this paper establishes the functional relationship between the measured value and the ideal value. Using the deviation between the ideal scribing position and the actual scribing position, we obtain the calibration curves. The calibration principle is deduced and analyzed in detail. The calibration angle measurement system is built and verified by experiments. The experimental results show that the original angle measurement error of the single readhead angle measurement sensor is 734.8", and the error after calibration is 2.4", and the calibration effect is better than that of the commonly used harmonic compensation method. The repeatability of the calibration system is better than 0.13". The in-situ calibration method can effectively reduce the angle measurement error, and the method is simple and the calibration efficiency is high. © 2019, Chinese Lasers Press. All right reserved.
引用
收藏
相关论文
共 14 条
[1]  
Wu Y.L., Wan H.J., A calibration algorithm of angle measuring error based on coordinate transformation, Electronic Science and Technology, 29, 9, (2016)
[2]  
Zhang L., Xiang Y., Correction algorithm of systemic errors in grating lateral-shearing interferometer, Chinese Journal of Lasers, 45, 8, (2018)
[3]  
Geckeler R.D., Just A., A shearing-based method for the simultaneous calibration of angle measuring devices, Measurement Science and Technology, 25, 10, (2014)
[4]  
Probst R., Self-calibration of divided circles on the basis of a prime factor algorithm, Measurement Science and Technology, 19, 1, (2008)
[5]  
Palmer E.W., Goniometer with continuously rotating gratings for use as an angle standard, Precision Engineer, 10, 3, pp. 147-152, (1988)
[6]  
Masuda T., Kajitani M., An automatic calibration system for angular encoders, Precision Engineering, 11, 2, pp. 95-100, (1989)
[7]  
Zhang J.J., Zhang Z.Y., Liu J.C., Novel method of calibrating the angle-measurement error of resolver, Chinese Journal of Scientific Instrument, 31, 1, pp. 149-153, (2010)
[8]  
Huang Y., Xue Z., He Y.W., Comparison between angle interferometer and angle encoder during calibration of autocollimator, Proceedings of SPIE, 9446, (2015)
[9]  
Zhai Z.S., Cheng Z., Zhang Y.H., Et al., Optical analysis of Moiré fringes of concentric-circle gratings, Laser & Optoelectronics Progress, 55, 7, (2018)
[10]  
Liu S., Installation and alignment of circular grating encoder, Equipment for Electronic Products Manufacturing, 47, 4, pp. 46-48, (2018)