Characterization of 3-dimensional printing and casting materials for use in magnetic resonance imaging phantoms at 3 t

被引:0
作者
Yunker B.E. [1 ,2 ]
Stupic K.F. [1 ]
Wagner J.L. [2 ]
Huddle S. [2 ]
Shandas R. [2 ]
Weir R.F. [2 ]
Russek S.E. [1 ]
Keenan K.E. [1 ]
机构
[1] Physical Measurement Laboratory, National Institute of Standards and Technology, Boulder, 80305, CO
[2] University of Colorado-Denver/Anschutz, Aurora, 80045, CO
关键词
3D printing; Medical imaging; MRI; Phantom; Polymer;
D O I
10.6028/JRES.125.028
中图分类号
学科分类号
摘要
Imaging phantoms are used to calibrate and validate the performance of magnetic resonance imaging (MRI) systems. Many new materials have been developed for additive manufacturing (three-dimensional [3D] printing) processes that may be useful in the direct printing or casting of dimensionally accurate, anatomically accurate, patient-specific, and/or biomimetic MRI phantoms. The T1, T2, and T2* spin relaxation times of polymer samples were tested to discover materials for use as tissue mimics and structures in MRI phantoms. This study included a cohort of polymer compounds that was tested in cured form. The cohort consisted of 101 standardized polymer samples fabricated from: two-part silicones and polyurethanes used in commercial casting processes; one-part optically cured polyurethanes used in 3D printing; and fused deposition thermoplastics used in 3D printing. The testing was performed at 3 T using inversion recovery, spin echo, and gradient echo sequences for T1, T2, and T2*, respectively. T1, T2, and T2* values were plotted with error bars to allow the reader to assess how well a polymer matches a tissue for a specific application. A correlation was performed between T1, T2, T2* values and material density, elongation, tensile strength, and hardness. Two silicones, SI_XP-643 and SI_P-45, may be usable mimics for reported liver values; one silicone, SI_XP-643, may be a useful mimic for muscle; one silicone, SI_XP-738, may be a useful mimic for white matter; and four silicones, SI_P-15, SI_GI-1000, SI_GI-1040, and SI_GI-1110, may be usable mimics for spinal cord. Elongation correlated to T2 (p = 0.0007), tensile strength correlated to T1 (p = 0.002), T2 (p = 0.0003), and T2* (p = 0.003). The 80 samples not providing measurable signal with T1, T2, T2* relaxation values too short to measure with the standard sequences, may be useful for MRI-invisible fixturing and medical devices at 3 T. © 2020 National Institute of Standards and Technology. All rights reserved.
引用
收藏
相关论文
共 43 条
[21]  
Mitsouras D, Lee TC, Liacouras P, Ionita CN, Pietilla T, Maier SE, Mulkern RV, Three-dimensional printing of MRI-visible phantoms and MR image-guided therapy simulation, Magnetic Resonance in Medicine, 77, 2, pp. 613-622, (2017)
[22]  
Yunker BE, Cordes D, Scherzinger AL, Dodd GD, Shandas R, Feng Y, Hunter KS, An investigation of industrial molding compounds for use in 3D ultrasound, MRI, and CT imaging phantoms, Medical Physics, 40, 5, (2013)
[23]  
Yunker BE, Dodd GD, Chen SJ, Chang S, Lanning CJ, Scherzinger AL, Shandas R, Feng Y, Hunter KS, The design and fabrication of two portal vein flow phantoms by different methods, Medical Physics, 41, 2, (2014)
[24]  
Russek SE, Stupic KF, Biller JR, Boss MA, Keenan KE, Mirowski E, Electromagnetics for Quantitative Magnetic Resonance Imaging, Compendium on Electromagnetic Analysis, pp. 95-147, (2020)
[25]  
Clique H, Cheng H-LM, Marie P-Y, Felblinger J, Beaumont M, 3D myocardial T1 mapping at 3T using variable flip angle method: Pilot study, Magnetic Resonance in Medicine, 71, 2, pp. 823-829, (2014)
[26]  
Preibisch C, Deichmann R, Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles, Magnetic Resonance in Medicine, 61, 1, pp. 125-135, (2009)
[27]  
Du J, Carl M, Bydder M, Takahashi A, Chung CB, Bydder GM, Qualitative and quantitative ultrashort echo time (UTE) imaging of cortical bone, Journal of Magnetic Resonance, 207, 2, pp. 304-311, (2010)
[28]  
Rakow-Penner R, Daniel B, Yu H, Sawyer-Glover A, Glover GH, Relaxation times of breast tissue at 1.5T and 3T measured using IDEAL, Journal of Magnetic Resonance Imaging, 23, 1, pp. 87-91, (2006)
[29]  
Bae WC, Chen PC, Chung CB, Masuda K, D'Lima D, Du J, Quantitative ultrashort echo time (UTE) MRI of human cortical bone: correlation with porosity and biomechanical properties, Journal of Bone and Mineral Research, 27, 4, pp. 848-857, (2012)
[30]  
Han E, Gold G, Stainsby J, Wright G, Beaulieu C, Brittain J, In-Vivo T1 and T2 Measurements of Muskuloskeletal Tissue at 3T and 1.5T, Proc ISMRM, 11, (2003)