Accuracy of oscillometric noninvasive blood pressure at the ankle in the lateral position during general anesthesia

被引:0
作者
Farag, Ahmed [1 ]
Mostafa, Maha [1 ]
Hasanin, Ahmed [1 ]
Sobh, Lamis [1 ]
Raafat, Heba [1 ]
Obayah, Gihan [1 ]
Youssef, Maha [1 ]
机构
[1] Cairo Univ, Fac Med, Dept Anesthesia & Crit Care Med, 01 Elsarayah St, Cairo 11559, Egypt
关键词
Oscillometry; Noninvasive blood pressure; Invasive blood pressure; Lateral position; Ankle; Hypotension; Hypertension; MEASURING DEVICES; RECOMMENDATIONS; VALIDATION; SOCIETY; CUFF;
D O I
10.1186/s12871-024-02595-6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background This study aimed to evaluate the accuracy of ankle blood pressure measurements in relation to invasive blood pressure in the lateral position. Methods This prospective observational study included adult patients scheduled for elective non-cardiac surgery under general anesthesia in the lateral position. Paired radial artery invasive and ankle noninvasive blood pressure readings were recorded in the lateral position using GE Carescape B650 monitor. The primary outcome was the ability of ankle mean arterial pressure (MAP) to detect hypotension (MAP < 70 mmHg) using area under the receiver operating characteristic curve (AUC) analysis. The secondary outcomes were the ability of ankle systolic blood pressure (SBP) to detect hypertension (SBP > 140 mmHg) as well as bias (invasive measurement - noninvasive measurement), and agreement between the two methods using the Bland-Altman analysis. Results We analyzed 415 paired readings from 30 patients. The AUC (95% confidence interval [CI]) of ankle MAP for detecting hypotension was 0.88 (0.83-0.93). An ankle MAP of <= 86 mmHg had negative and positive predictive values (95% CI) of 99 (97-100)% and 21 (15-29)%, respectively, for detecting hypotension. The AUC (95% CI) of ankle SBP to detect hypertension was 0.83 (0.79-0.86) with negative and positive predictive values (95% CI) of 95 (92-97)% and 36 (26-46)%, respectively, at a cutoff value of > 144 mmHg. The mean bias between the two methods was - 12 +/- 17, 3 +/- 12, and - 1 +/- 11 mmHg for the SBP, diastolic blood pressure, and MAP, respectively. Conclusion In patients under general anesthesia in the lateral position, ankle blood pressure measurements are not interchangeable with the corresponding invasive measurements. However, an ankle MAP > 86 mmHg can exclude hypotension with 99% accuracy, and an ankle SBP < 144 mmHg can exclude hypertension with 95% accuracy.
引用
收藏
页数:6
相关论文
共 13 条
[1]   Oscillometric blood pressure: a review for clinicians [J].
Alpert, Bruce S. ;
Quinn, David ;
Gallick, David .
JOURNAL OF THE AMERICAN SOCIETY OF HYPERTENSION, 2014, 8 (12) :930-938
[2]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[3]   AAGBI recommendations for standards of monitoring during anaesthesia and recovery 2015 [J].
Checketts, M. R. .
ANAESTHESIA, 2016, 71 (04) :470-471
[4]   A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output [J].
Critchley, Lester A. ;
Lee, Anna ;
Ho, Anthony M. -H. .
ANESTHESIA AND ANALGESIA, 2010, 111 (05) :1180-1192
[5]   Evaluation of the accuracy of oscillometric non-invasive blood pressure measurement at the ankle in children during general anesthesia [J].
Gamal, Reham M. ;
Mostafa, Maha ;
Hasanin, Ahmed M. ;
Khedr, Suzan Adlan ;
Abdelgalil, Ahmed Salah ;
Elshal, Mamdouh Mahmoud .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2023, 37 (05) :1239-1245
[6]   Noninvasive Monitoring of Arterial Pressure: Finger or Lower Leg As Alternatives to the Upper Arm: A Prospective Study in Three ICUs [J].
Lakhal, Karim ;
Dauvergne, Jerome E. ;
Kamel, Toufik ;
Messet-Charriere, Helene ;
Jacquier, Sophie ;
Robert-Edan, Vincent ;
Nay, Mai-Anh ;
Rozec, Bertrand ;
Ehrmann, Stephan ;
Mueller, Gregoire ;
Boulain, Thierry .
CRITICAL CARE MEDICINE, 2023, 51 (10) :1350-1362
[7]   Noninvasive monitoring of blood pressure in the critically ill: Reliability according to the cuff site (arm, thigh, or ankle) [J].
Lakhal, Karim ;
Macq, Christine ;
Ehrmann, Stephan ;
Boulain, Thierry ;
Capdevila, Xavier .
CRITICAL CARE MEDICINE, 2012, 40 (04) :1207-1213
[8]   Accuracy of oscillometric blood pressure measurement at both arms in the lateral position [J].
Mostafa, Maha ;
Hasanin, Ahmed M. ;
Elsayed, Omnia S. ;
Mostafa, Mamdouh M. ;
Sarhan, Khaled .
BLOOD PRESSURE MONITORING, 2021, 26 (05) :364-372
[9]   European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults [J].
O'Brien, Eoin ;
Atkins, Neil ;
Stergiou, George ;
Karpettas, Nikos ;
Parati, Gianfranco ;
Asmar, Roland ;
Imai, Yutaka ;
Wang, Jiguang ;
Mengden, Thomas ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :23-38
[10]   Intraoperative Blood Pressure Monitoring in Obese Patients Arterial Catheter, Finger Cuff, and Oscillometry [J].
Schumann, Roman ;
Meidert, Agnes S. ;
Bonney, Iwona ;
Koutentis, Christos ;
Wesselink, Wilbert ;
Kouz, Karim ;
Saugel, Bernd .
ANESTHESIOLOGY, 2021, 134 (02) :179-188