Feasibility of patient-specific quality assurance (PSQA) for real-time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces

被引:0
|
作者
Xu, Qianyi [1 ,2 ]
Fan, Jiajin [1 ]
Vinogradskiy, Yevgeniy [2 ]
Chawla, Ashish K. [1 ]
Kubicek, Gregory [3 ]
Yang, Haihua [4 ]
Huynh, Kiet [1 ]
LaCouture, Tamara [2 ]
Grimm, Jimm [5 ]
Nie, Wei [1 ]
机构
[1] Inova Schar Canc Inst, Dept Adv Radiat Oncol & Proton Therapy, Fairfax, VA USA
[2] Thomas Jefferson Univ, Dept Radiat Oncol, 111 South 11th St, Philadelphia, PA 19107 USA
[3] Univ Miami, Dept Radiat Oncol, Miami, FL USA
[4] Taizhou Hosp, Dept Radiat Oncol, Taizhou, Zhejiang, Peoples R China
[5] Dept Radiat Oncol, Wellstar Hlth Syst, Marietta, GA USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2024年 / 25卷 / 08期
关键词
CyberKnife; diode array; motion platform; PSQA; synchrony; RADIATION-THERAPY; TRACKING SYSTEM; ACCURACY; SRS;
D O I
10.1002/acm2.14352
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation. Methods: The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior-inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0 degrees +/- 0.15 degrees or 1.2 degrees +/- 0.1 degrees) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom. Results: For smaller pitch angles, the mean gamma passing rates were 99.75% +/- 0.87%, 98.63% +/- 2.05%, and 93.11% +/- 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% +/- 10.91%, 88.05% +/- 14.93%, and 80.38% +/- 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles (p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles (p < 0.02 for all criteria). All the gamma passing rates were improved, if applying shift and rotation correction. Conclusions: The significance of this work is that it is the first study to benchmark PSQA for the CyberKnife Synchrony system using realistically moving phantoms with rotation. With reasonable delivery time, we found it may be feasible to perform PSQA for Synchrony patients with a realistic breathing pattern.
引用
收藏
页数:9
相关论文
共 30 条
  • [1] A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy
    Shiinoki, Takehiro
    Fujii, Fumitake
    Fujimoto, Koya
    Yuasa, Yuki
    Sera, Tatsuhiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07): : 16 - 28
  • [2] Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy
    Xu, Qianyi
    Huynh, Kiet
    Nie, Wei
    Rose, Mark S.
    Chawla, Ashish K.
    Choe, Kevin S.
    Kanani, Samir
    Kubicek, Gregory J.
    Fan, Jiajin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [3] Matched linac stereotactic radiotherapy: An assessment of delivery similarity and distributive patient-specific quality assurance feasibility
    Goodall, Simon K.
    Dunn, Leon
    Dunning, Jonathan
    Munoz, Luis
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (11):
  • [4] Evaluation of a novel patient-specific quality assurance phantom for robotic single-isocentre, multiple-target stereotactic radiosurgery, and stereotactic radiotherapy
    Xu, Qianyi
    Vinogradskiy, Yevgeniy
    Grimm, Jimm
    Nie, Wei
    Dupre, Pamela
    Chawla, Ashish K.
    Bajaj, Gopal
    Yang, Haihua
    Lacouture, Tamara
    Fan, Jiajin
    BRITISH JOURNAL OF RADIOLOGY, 2024, 97 (1155) : 660 - 667
  • [5] A six-degree-of-freedom robotic motion system for quality assurance of real-time image-guided radiotherapy
    Alnaghy, Saree
    Kyme, Andre
    Caillet, Vincent
    Doan Trang Nguyen
    O'Brien, Ricky
    Booth, Jeremy T.
    Keall, Paul J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10)
  • [6] Real-time intrafraction prostate motion during linac based stereotactic radiotherapy with rectal displacement
    Legge, Kimberley
    Doan Nguyen
    Ng, Jin Aun
    Wilton, Lee
    Richardson, Matthew
    Booth, Jeremy
    Keall, Paul
    O'Connor, Darryl J.
    Greer, Peter
    Martin, Jarad
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (06): : 130 - 136
  • [7] respiTrack: Patient-specific real-time respiratory tumor motion prediction using magnetic tracking
    Ozbek, Yusuf
    Bardosi, Zoltan
    Freysinger, Wolfgang
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2020, 15 (06) : 953 - 962
  • [8] Real-time measurement of ICD lead motion during stereotactic body radiotherapy of ventricular tachycardia
    Knybel, Lukas
    Cvek, Jakub
    Neuwirth, Radek
    Jiravsky, Otakar
    Hecko, Jan
    Penhaker, Marek
    Sramko, Marek
    Kautzner, Josef
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2021, 26 (01) : 128 - 137
  • [9] Assessing Feasibility of Real-time Ultrasound Monitoring in Stereotactic Body Radiotherapy of Liver Tumors
    Zhong, Yahua
    Stephans, Kevin
    Qi, Peng
    Yu, Naichang
    Wong, John
    Xia, Ping
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2013, 12 (03) : 243 - 250
  • [10] Clinical implementation and patient-specific quality assurance solutions for real-time target tracking and dynamic delivery in Radixact synchrony
    Trujillo-Bastidas, Christian D.
    Taylor, Michael J.
    Diaz-Londono, Gloria M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):