Linearizing the vertical scale of an interferometric microscope and its effect on step-height measurement

被引:0
作者
Germer, Thomas A. [1 ]
Renegar, T. Brian [1 ]
Griesmann, Ulf [1 ]
Soons, Johannes A. [1 ]
机构
[1] Natl Inst Stand & Technol, Sensor Sci Div, Gaithersburg, MD 20899 USA
关键词
microscope; linearization; interferometer; calibration; step-height; coherence scanning interferometer; CALIBRATION;
D O I
10.1088/2051-672X/ad44bb
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The vertical scale calibration of an interferometric microscope is important for establishing traceability of surface topography measurements to the International System of Units (SI) unit of length, the meter. Building on the calibration procedure for the amplification coefficient developed by de Groot and Beverage [Proc. SPIE 9526, 952610 (2015)], this paper describes a calibration procedure that yields the response curve for the entire vertical scan motion of a coherent scanning interferometric microscope. The method requires only a flat mirror as an artifact, a narrow band spectral filter, an aperture to reduce the effective numerical aperture, and the ability to raise and lower the microscope head so that the center of the interferogram can be varied within the scan range. The local frequency of the interferogram is determined by fitting sections of the interferogram to a sinusoidal function. The nonlinearity determined from the local frequency data can be used to estimate the uncertainty in uncorrected vertical height measurements. We describe how optical profile data can be corrected for nonlinearity due to dynamic effects in the scan motion and show that the correction improves the reproducibility of step height measurements by at least a factor of three and close to that of the repeatability.
引用
收藏
页数:11
相关论文
共 21 条
[1]   Simple method to determine linearity deviations of topography measuring instruments with a large range axial scanning system [J].
Bauer, Wilfried ;
Hueser, Dorothee ;
Gerbert, David .
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2020, 64 :243-248
[2]   Calibration of the z-axis for large-scale scanning white-light interferometers [J].
Boedecker, S. ;
Rembe, C. ;
Schmid, H. ;
Hageney, T. ;
Koehnlein, T. .
13TH INTERNATIONAL CONFERENCE ON METROLOGY AND PROPERTIES OF ENGINEERING SURFACES, 2011, 311
[3]   CALIBRATION OF NUMERICAL APERTURE EFFECTS IN INTERFEROMETRIC MICROSCOPE OBJECTIVES [J].
CREATH, K .
APPLIED OPTICS, 1989, 28 (16) :3333-3338
[4]   Does interferometry work? A critical look at the foundations of interferometric surface topography measurement [J].
de Groot, Peter ;
de Lega, Xavier Colonna ;
Su, Rong ;
Leach, Richard .
APPLIED OPTICAL METROLOGY III, 2019, 11102
[5]   Calibration of the amplification coefficient in interference microscopy by means of a wavelength standard [J].
de Groot, Peter ;
Beverage, Jake .
MODELING ASPECTS IN OPTICAL METROLOGY V, 2015, 9526
[6]   Principles of interference microscopy for the measurement of surface topography [J].
de Groot, Peter .
ADVANCES IN OPTICS AND PHOTONICS, 2015, 7 (01) :1-65
[7]  
de Groot P, 2011, OPTICAL MEASUREMENT OF SURFACE TOPOGRAPHY, P187
[8]  
de Groot P, 2011, OPTICAL MEASUREMENT OF SURFACE TOPOGRAPHY, P167
[9]   Manufacturing of new roughness standards for the linearity of the vertical axis - Feasibility study and optimization [J].
Eifler, Matthias ;
Schneider, Frank ;
Seewig, Joerg ;
Kirsch, Benjamin ;
Aurich, Jan C. .
ENGINEERING SCIENCE AND TECHNOLOGY-AN INTERNATIONAL JOURNAL-JESTECH, 2016, 19 (04) :1993-2001
[10]   Improved and simpler estimation of scale linearity contribution to topography measurement [J].
Giusca, Claudiu L. ;
Goel, Saurav .
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2019, 60 :368-373