Three-dimensional finite element analysis of the biomechanical properties of different material implants for replacing missing teeth

被引:0
作者
Gao, Yichen [1 ,2 ]
He, Xianyi [1 ,2 ]
Xu, Wei [1 ,2 ]
Deng, Yuyao [1 ,2 ]
Xia, Zhaoxin [1 ,2 ]
Chen, Junliang [1 ,2 ]
He, Yun [1 ,2 ]
机构
[1] Southwest Med Univ, Affiliated Stomatol Hosp, Dept Oral & Maxillofacial Surg, Luzhou 646000, Peoples R China
[2] Southwest Med Univ, Affiliated Stomatol Hosp, Oral & Maxillofacial Reconstruct & Regenerat Luzho, Luzhou 646000, Peoples R China
关键词
Finite element analysis; Zirconia; Titanium; Titanium-zirconia; Poly(ether-ether-ketone); DENTAL IMPLANTS; STRESS-DISTRIBUTION; BONE; TITANIUM; DESIGN; OSSEOINTEGRATION; PEEK;
D O I
10.1007/s10266-024-00942-0
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to analyze the biomechanical properties of implants made of different materials to replace missing teeth by using three-dimensional finite element analysis and provide a theoretic basis for clinical application. CBCT data was imported into the Mimics and 3-Matic to construct the three-dimensional finite element model of a missing tooth restored by an implant. Then, the model was imported into the Marc Mentat. Based on the variations of the implant materials (titanium, titanium-zirconia, zirconia and poly (ether-ether-ketone) (PEEK)) and bone densities (high and low), a total of eight models were created. An axial load of 150 N was applied to the crown of the implant to simulate the actual occlusal situation. Both the maximum values of stresses in the cortical bone and implant were observed in the Zr-low model. The maximum displacements of the implants were also within the normal range except for the PEEK models. The cancellous bone strains were mainly distributed in the apical area of the implant, and the maximum value (3225 mu strain) was found in PEEK-low model. Under the premise of the same implant material, the relevant data from various indices in low-density bone models were larger than that in high-density bone models. From the biomechanical point of view, zirconia, titanium and titanium-zirconia were all acceptable implant materials for replacing missing teeth and possessed excellent mechanical properties, while the application of PEEK material needs to be further optimized and modified.
引用
收藏
页码:80 / 88
页数:9
相关论文
共 60 条
[21]   Bone stability around dental implants: Treatment related factors [J].
Heinemann, Friedhelm ;
Hasan, Istabrak ;
Bourauel, Christoph ;
Biffar, Reiner ;
Mundt, Torsten .
ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2015, 199 :3-8
[22]   A Finite Element Analysis to Compare Stress Distribution on Extra-Short Implants with Two Different Internal Connections [J].
Helena Garcia-Braz, Silvia ;
Prados-Privado, Maria ;
Silveira Zanatta, Luiz Carlos ;
Luis Calvo-Guirado, Jose ;
Carlos Prados-Frutos, Juan ;
Alexandre Gehrke, Sergio .
JOURNAL OF CLINICAL MEDICINE, 2019, 8 (08)
[23]   Three-Dimensional Finite Element Analysis of Worn Molars With Prosthetic Crowns and Onlays Made of Various Materials [J].
Houdaifa, Raafat ;
Alzoubi, Hasan ;
Jamous, Issam .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (10)
[24]   The effect of implant neck microthread design on stress distribution of peri-implant bone with different level: A finite element analysis [J].
Jin, Zhi-Heng ;
Peng, Meng-Dong ;
Li, Qing .
JOURNAL OF DENTAL SCIENCES, 2020, 15 (04) :466-471
[25]  
John Jins, 2012, J Indian Prosthodont Soc, V12, P37, DOI 10.1007/s13191-012-0114-0
[26]   Cortical bone thickness in dentate and edentulous human cadavers [J].
Katranji, Amar ;
Misch, Kelly ;
Wang, Hom-Lay .
JOURNAL OF PERIODONTOLOGY, 2007, 78 (05) :874-878
[27]   MECHANICAL-PROPERTIES OF THE BINARY TITANIUM-ZIRCONIUM ALLOYS AND THEIR POTENTIAL FOR BIOMEDICAL MATERIALS [J].
KOBAYASHI, E ;
MATSUMOTO, S ;
DOI, H ;
YONEYAMA, T ;
HAMANAKA, H .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (08) :943-950
[28]   Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog [J].
Koch, F. P. ;
Weng, D. ;
Kraemer, S. ;
Biesterfeld, S. ;
Jahn-Eimermacher, A. ;
Wagner, W. .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) :350-356
[29]   The Etiology and Management of Dental Implant Failure: A Review [J].
Kochar, Shraddha P. ;
Reche, Amit ;
Paul, Priyanka .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (10)
[30]   Effectiveness and Mechanisms of Low-Intensity Pulsed Ultrasound on Osseointegration of Dental Implants and Biological Functions of Bone Marrow Mesenchymal Stem Cells [J].
Liang, Chao ;
Liu, Xiu ;
Yan, Yuwei ;
Sun, Rongxin ;
Li, Jun ;
Geng, Wei .
STEM CELLS INTERNATIONAL, 2022, 2022