A Comparative Analysis of Freelite and N Latex FLC Assay for the sFLC Measurements in Suspected Patients of Monoclonal Gammopathy

被引:1
作者
Singh, Manish K. [1 ]
Gupta, Priyanka [1 ]
Kumar, Kaushal [1 ]
Chandra, Dinesh [1 ]
Rahman, Khaliqur [1 ]
Gupta, Ruchi [1 ]
Yadav, Sanjeev [1 ]
Kashyap, Rajesh [1 ]
机构
[1] SGPGIMS, Dept Hematol, Lucknow, India
关键词
sFLC; Polyclonal; Monoclonal; N Latex assay; M-protein; FREE LIGHT-CHAINS; SERUM; KAPPA; PERFORMANCE; CRITERIA; LAMBDA;
D O I
10.1007/s12288-024-01747-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
With the introduction of newer platforms for sFLC estimation, more comparative attempts are required to establish an assay-specific reference range, that can be used in the day-to-day clinical management of the patients, as the revised criteria for the diagnosis of myeloma recommend the use of involved Free light chain (FLC)/uninvolved FLC ratio as myeloma defining event, solely relying on Freelite assay. This was an observational study that included 101 patients, 55 (54.5%) with M-protein (myeloma), and 46 (45.5%) without M-protein. The estimation of the sFLC assay was carried out at two separate platforms, the Freelite Assay (The binding Site, Birmingham, UK) on a Siemens BNII Nephelometer, and the N Latex FLC reagent on Atellica NEPH 630 (Siemens Healthineers, Eschborn, Germany). The comparative correlational analysis of the two assays for estimation of kappa-, lambda FLC, and kappa/lambda ratio showed a strong to a very strong association (P-value < 0.005) among total patients, however, this was reduced to weak to moderate association among patients with M-protein (P-value < 0.005). The contingency analyses to evaluate the agreement between the assays also showed a similar trend. A better concordance for the kappa/lambda ratio between the two assays was achieved while expanding this ratio. Our study concluded that two assays showed a better agreement among the patients without M-protein, but these two cannot be used interchangeably for the clinical management of myeloma patients.
引用
收藏
页码:694 / 700
页数:7
相关论文
共 20 条
[1]  
Altman D.G., 1990, PRACTICAL STAT MED R, P404
[2]   Diagnostic thresholds for free light chains in multiple myeloma depend on the assay used [J].
Bossuyt, Xavier ;
Delforge, Michel ;
Reynders, Martin ;
Dillaerts, Doreen ;
Sprangers, Ben ;
Fostier, Karel ;
Poesen, Koen ;
Vercammen, Martine .
LEUKEMIA, 2018, 32 (08) :1815-1818
[3]  
Bradwell AR, 2001, CLIN CHEM, V47, P673
[4]   Development of a rapid and quantitative lateral flow assay for the simultaneous measurement of serum κ and λ immunoglobulin free light chains (FLC): inception of a new near-patient FLC screening tool [J].
Campbella, John P. ;
Heaneya, Jennifer L. J. ;
Shemar, Meena ;
Baldwin, Dene ;
Griffin, Ann E. ;
Oldridge, Emma ;
Goodall, Margaret ;
Afzal, Zaheer ;
Plant, Tim ;
Cobbold, Mark ;
Jefferis, Roy ;
Jacobs, Joannes F. M. ;
Hand, Christopher ;
Drayson, Mark T. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (03) :424-434
[5]   Analytical issues of serum free light chain assays and the relative performance of polyclonal and monoclonal based reagents [J].
Carr-Smith, Hugh D. ;
Jenner, Ellen L. ;
Evans, Josie A. R. ;
Harding, Stephen J. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (06) :997-1003
[6]   Polyclonal versus monoclonal immunoglobulin-free light chains quantification [J].
Di Noto, Giuseppe ;
Cimpoies, Elena ;
Dossi, Alessandra ;
Paolini, Lucia ;
Radeghieri, Annalisa ;
Caimi, Luigi ;
Ricotta, Doris .
ANNALS OF CLINICAL BIOCHEMISTRY, 2015, 52 (03) :327-336
[7]  
Evans JD, 1996, STRAIGHTFORWARD STAT, DOI DOI 10.2307/2291607
[8]   Evaluation of the N-latex serum free light chain assay on the Siemens BNII analyzer and agreement with The Binding Site FreeLite assay on the SPAPlus [J].
Habeeb, Nicole M. A. White-Al ;
Earle, Tammy ;
Spencer, Megan ;
Blasutig, Ivan M. .
CLINICAL BIOCHEMISTRY, 2018, 51 :90-96
[9]   Prognostic value of involved/uninvolved free light chain ratio determined by Freelite and N Latex FLC assays for identification of high-risk smoldering myeloma patients [J].
Henriot, Basile ;
Rouger, Emmanuel ;
Rousseau, Chloe ;
Escoffre, Martine ;
Sebillot, Martine ;
Bendavid, Claude ;
Minvielle, Stephane ;
Avet-Loiseau, Herve ;
Decaux, Olivier ;
Moreau, Caroline .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (09) :1397-1405
[10]   Clinical comparison of new monoclonal antibody-based nephelometric assays for free light chain kappa and lambda to polyclonal antibody-based assays and immunofixation electrophoresis [J].
Hoedemakers, Rein M. J. ;
Pruijt, Johannes F. M. ;
Hol, Suzanne ;
Teunissen, Elisa ;
Martens, Henk ;
Stam, Peter ;
Melsert, Roel ;
Velthuis, Henk Te .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (03) :489-495