EVALUATION OF LATEX AGGLUTINATION AND MICROTUBE COAGULASE TESTS FOR DETECTION OF STAPHYLOCOCCUS-AUREUS

被引:7
|
作者
POURSHADI, M
KLAAS, J
机构
关键词
D O I
10.1016/0732-8893(84)90059-2
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
In a blind study, a latex agglutination test (Serostat Staphylococcus) and a microtube coagulase test (Staphase) were evaluated for their ability to detect S. aureus. Of 289 isolates of catalase-positive, gram-positive cocci, 122 were identified as S. aureus based on positive reactions in at least 3 of the following tests: tube coagulase, slide coagulase, DNase production or anaerobic fermentation of mannitol. The latex agglutination test gave positive reactions for all S. aureus isolates and 10 (6%) non-S. aureus isolates. The slide coagulase test was positive for 121 S. aureus isolates and 3 (2%) non-S. aureus isolates. The microtube coagulase test detected 53, 90 and 98% of the S. aureus isolates after 2, 4, and 24 h, respectively. The conventional tube coagulase test detected 97% of the S. aureus isolates after 2 h, and 98% after 4 and 24 h. Two isolates of S. aureus gave negative tube coagulase reactions at 37.degree. C, but positive reactions at room temperature after 24 h. The combination of tube and slide coagulase tests provided the most reliable results. The slide and tube coagulase tests gave more reliable results than the latex agglutination and microtube coagulase tests, respectively.