CONTROLLED EVALUATION OF BACT ALERT STANDARD ANAEROBIC AND FAN ANAEROBIC BLOOD CULTURE BOTTLES FOR THE DETECTION OF BACTEREMIA AND FUNGEMIA

被引:62
|
作者
WILSON, ML
WEINSTEIN, MP
MIRRETT, S
REIMER, LG
FELDMAN, RJ
CHUARD, CR
RELLER, LB
机构
[1] DENVER HLTH & HOSP,DEPT PATHOL & LAB SERV,DENVER,CO 80204
[2] UNIV COLORADO,SCH MED,DEPT PATHOL,DENVER,CO 80262
[3] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT MED,NEW BRUNSWICK,NJ 08901
[4] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT PATHOL,NEW BRUNSWICK,NJ 08901
[5] ROBERT WOOD JOHNSON UNIV HOSP,MICROBIOL LAB,NEW BRUNSWICK,NJ 08901
[6] DUKE UNIV,MED CTR,CLIN MICROBIOL LAB,DURHAM,NC 27710
[7] DUKE UNIV,MED CTR,DEPT PATHOL,DURHAM,NC 27710
[8] DUKE UNIV,MED CTR,DEPT MED,DURHAM,NC 27710
[9] DEPT VET AFFAIRS MED CTR,MICROBIOL LAB,SALT LAKE CITY,UT 84148
[10] UNIV UTAH,SCH MED,DEPT MED,SALT LAKE CITY,UT 84148
[11] UNIV UTAH,SCH MED,DEPT PATHOL,SALT LAKE CITY,UT 84148
关键词
D O I
10.1128/JCM.33.9.2265-2270.1995
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
FAN medium was formulated to improve microbial recovery, particularly for fastidious microorganisms and for microorganisms causing sepsis in patients receiving antimicrobial therapy, In a controlled clinical evaluation performed at four university-affiliated hospitals, FAN anaerobic bottles were compared with standard anaerobic bottles for yield, speed of detection of microbial growth, and detection of septic episodes, A total of 10,431 blood culture sets were received; both anaerobic bottles of 7,694 blood culture sets were adequately filled with blood, Altogether, 925 isolates were recovered: 557 that were the cause of sepsis, 99 that were indeterminate as the cause of sepsis, and 269 contaminants, More Staphylococcus aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Escherichia coli (P < 0.02), and all microorganisms combined (P < 0.005) were recovered from FAN bottles; more nonfermentative gram-negative bacilli (P < 0.05), Torulopsis glabrata (P < 0.001), and other yeasts (P < 0.01) were recovered from standard bottles, Growth of S. aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Enterococcus faecalis (P < 0.025), streptococci other than Streptococcus pneumoniae (P < 0.01), and all microorganisms combined (P < 0.001) was detected earlier in standard bottles; growth of more isolates of E. coli (P < 0.05) and anaerobic bacteria (P < 0.01) was detected earlier in FAN bottles. The mean times to detection were 14.2 and 16.1 h for standard and FAN bottles, respectively, More septic episodes caused by S. aureus (P < 0.001), coagulase-negative staphylococci (P < 0.005), members of the family Enterobacteriaceae (P < 0.02), and all microorganisms combined (P < 0.02) were detected in FAN bottles; more septic episodes caused by nonfermentative gram-negative bacilli (P < 0.025) and yeasts (P < 0.005) were detected in standard bottles, In summary, more isolates (except for strict aerobes) were recovered from FAN bottles than from standard anaerobic bottles, Similarly, significantly more septic episodes (except for those caused by strict aerobes) were detected with FAN bottles than with standard anaerobic bottles, With the exception of E. coli and anaerobic bacteria, growth of more isolates was detected earlier in standard anaerobic bottles.
引用
收藏
页码:2265 / 2270
页数:6
相关论文
共 50 条
  • [41] Anaerobic Bottles are Still Important in Blood Culture Sets
    P. Khanna
    P. Collignon
    European Journal of Clinical Microbiology and Infectious Diseases, 2001, 20 : 217 - 219
  • [42] Time to Detection with BacT/Alert FA Plus Compared to BacT/Alert FA Blood Culture Media
    Nutman, A.
    Even-Tsur, S. Fisher
    Shapiro, G.
    Braun, T.
    Schwartz, D.
    Carmeli, Y.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2016, 35 (09) : 1469 - 1473
  • [43] Time to Detection with BacT/Alert FA Plus Compared to BacT/Alert FA Blood Culture Media
    A. Nutman
    S. Fisher Even-Tsur
    G. Shapiro
    T. Braun
    D. Schwartz
    Y. Carmeli
    European Journal of Clinical Microbiology & Infectious Diseases, 2016, 35 : 1469 - 1473
  • [44] Direct detection of mecA, nuc and 16S rRNA genes in BacT/Alert blood culture bottles
    Lem, P
    Spiegelman, J
    Toye, B
    Ramotar, K
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2001, 41 (03) : 165 - 168
  • [45] In Vitro Evaluation of BacT/Alert FA Blood Culture Bottles and T2Candida Assay for Detection of Candida in the Presence of Antifungals
    Beyda, Nicholas D.
    Amadio, Jonathan
    Rodriguez, Jose R.
    Malinowski, Karen
    Garey, Kevin W.
    Wanger, Audrey
    Ostrosky-Zeichner, Luis
    JOURNAL OF CLINICAL MICROBIOLOGY, 2018, 56 (08)
  • [46] CONTROLLED EVALUATION OF BACTEC PLUS-27 AND ROCHE SEPTI-CHEK ANAEROBIC BLOOD CULTURE BOTTLES
    WILSON, ML
    HARRELL, LJ
    MIRRETT, S
    WEINSTEIN, MP
    STRATTON, CW
    RELLER, LB
    JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (01) : 63 - 66
  • [47] Clinical evaluation of FAPlus/FNPlus bottles compared with the combination of SA/SN and FA/FN bottles in the BacT/Alert blood culture system
    Hattori, Takuya
    Nishiyama, Hideki
    Ikegami, Shinobu
    Minoshima, Makoto
    Kato, Hideki
    Yuasa, Norihiro
    JOURNAL OF MEDICAL INVESTIGATION, 2020, 67 (1-2): : 90 - 94
  • [48] Inactivation of Clinically Frequently Used Antimicrobial Agents By BacT/ALERT FAN Plus and BACTEC Aerobic and Anaerobic Culture Media in Simulated Blood Cultures: First Comparative Evaluation in China Mainland
    Xu, Xiuyu
    Wei, Qiang
    Wang, Zhu
    Yan, Jie
    Wang, Huijuan
    Xia, Yun
    INFECTION AND DRUG RESISTANCE, 2021, 14 : 163 - 172
  • [49] Multicenter clinical comparison of Resin-containing bottles with standard aerobic and anaerobic bottles for culture of microorganisms from blood
    Lelievre, H
    Gimenez, M
    Vandenesch, F
    Reinhardt, A
    Lenhardt, D
    Just, HM
    Pau, M
    Ausina, V
    Etienne, J
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1997, 16 (09) : 669 - 674
  • [50] DETECTION OF BACTEREMIA IN VENTED AND UNVENTED BLOOD CULTURE BOTTLES
    HALL, M
    WARREN, E
    WASHINGTON, JA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1976, 65 (02) : 260 - 261