ARGON XENON PLASMA FOR ALLEVIATING POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:50
作者
SMITH, FG [1 ]
WIEDERIN, DR [1 ]
HOUK, RS [1 ]
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,AMES LAB,DEPT CHEM,AMES,IA 50011
关键词
D O I
10.1021/ac00014a021
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Xenon is added at 10 or 37 mL min-1 to the aerosol gas flow of an argon inductively coupled plasma mass spectrometer. Addition of xenon substantially reduces polyatomic ions such as N2+, HN2+, NO+, ArH+, CIO+, ArC+, CIOH+, ArN+, and ArO+ and facilitates the measurement of Si-28, Si-29, K-39, K-41, V-51, Cr-52, Cr-53, Cr-54, Fe-54, and Fe-58. Isotope ratios are determined with relative standard deviations from 0.6% to 1.6%. Detection limits for the above elements range from 0.3 to 2-mu-g L-1.
引用
收藏
页码:1458 / 1462
页数:5
相关论文
共 32 条