OCULAR HAZARD OF METALLIC FRAGMENTS DURING MR IMAGING AT 0.06 T

被引:13
|
作者
KELSEY, CA
KING, JN
KECK, GM
CHIU, MT
WOLFE, DM
ORRISON, WW
机构
[1] UNIV NEW MEXICO,SCH MED,DEPT NEUROL,ALBUQUERQUE,NM 87131
[2] UNIV NEW MEXICO,SCH MED,DEPT PHYS & ASTRON,ALBUQUERQUE,NM 87131
[3] EYE ASSOC NEW MEXICO,ALBUQUERQUE,NM
[4] UNIV NEW MEXICO HOSP,OPHTHALMOL CLIN,ALBUQUERQUE,NM
关键词
EYE; MR STUDIES; FOREIGN BODIES; MAGNETIC RESONANCE (MR); SAFETY;
D O I
10.1148/radiology.180.1.2052711
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The hazard of magnetic resonance (MR) imaging at low field strength (0.06 T) was examined by using both in vitro and in vivo animal experiments. Ferromagnetic fragments, 0.25-2.2 mm in diameter, were placed in vitreous humor obtained from excised cow eyes and in rabbit eyes. The magnetic field strength at which motion occurred was measured for each fragment size. Two types of motion, rotation and translation, were observed. The authors found that fragments that could not be seen on plain radiographs did not appear to move enough to be a significant ocular hazard during MR imaging at a field strength of 0.06 T.
引用
收藏
页码:282 / 283
页数:2
相关论文
共 50 条
  • [1] METALLIC BALLISTIC FRAGMENTS - MR IMAGING SAFETY AND ARTIFACTS
    TEITELBAUM, GP
    YEE, CA
    VANHORN, DD
    KIM, HS
    COLLETTI, PM
    RADIOLOGY, 1990, 175 (03) : 855 - 859
  • [2] WALLSTENT METALLIC BILIARY ENDOPROSTHESIS - MR IMAGING CHARACTERISTICS
    GIRARD, MJ
    HAHN, PF
    SAINI, S
    DAWSON, SL
    GOLDBERG, MA
    MUELLER, PR
    RADIOLOGY, 1992, 184 (03) : 874 - 876
  • [3] EXTRUSION OF EYE SOCKET MAGNETIC IMPLANT AFTER MR IMAGING - POTENTIAL HAZARD TO PATIENT WITH EYE PROSTHESIS
    YUH, WTC
    HANIGAN, MT
    NERAD, JA
    EHRHARDT, JC
    CARTER, KD
    KARDON, RH
    SHELLOCK, FG
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06): : 711 - 713
  • [4] Potential MR hazard to patients with metallic heart valves: The Lenz effect
    Condon, B
    Hadley, DM
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 12 (01) : 171 - 176
  • [5] PATIENT MONITORING DURING CLINICAL MR IMAGING
    KANAL, E
    SHELLOCK, FG
    RADIOLOGY, 1992, 185 (03) : 623 - 629
  • [6] MR imaging in the presence of small circular metallic implants - Assessment of thermal injuries
    Manner, I
    Alanen, A
    Komu, M
    Savunen, T
    Kantonen, I
    Ekfors, T
    ACTA RADIOLOGICA, 1996, 37 (04) : 551 - 554
  • [7] FAST IMAGING MR ASSESSMENT OF URETERONEPHROSIS DURING PREGNANCY
    ROY, C
    SAUSSINE, C
    JAHN, C
    LEBRAS, Y
    STEICHEN, G
    DELEPAUL, B
    CAMPOS, M
    CHAMBRON, J
    JACQMIN, D
    MAGNETIC RESONANCE IMAGING, 1995, 13 (06) : 767 - 772
  • [8] Aneurysm clips: Evaluation of MR imaging artifacts at 1.5 T
    Shellock, FG
    Kanal, E
    RADIOLOGY, 1998, 209 (02) : 563 - 566
  • [9] Interventional MR imaging at 1.5 T: Quantification of sound exposure
    Moelker, A
    Maas, RAJJ
    Lethimonnier, F
    Pattynama, PMT
    RADIOLOGY, 2002, 224 (03) : 889 - 895
  • [10] Assessment of Sensations Experienced by Subjects during MR Imaging Examination at 7T
    Uwano, Ikuko
    Metki, Tsuyoshi
    Sendai, Fusako
    Yoshida, Ryoko
    Kudo, Kohsuke
    Yamashita, Fumio
    Higuchi, Satomi
    Ito, Kenji
    Harada, Taisuke
    Goodwin, Jonathan
    Ogawa, Akira
    Sasaki, Makoto
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2015, 14 (01) : 35 - 41