CLINICAL OBSERVATIONS WITH A DUAL SENSOR RATE ADAPTIVE SINGLE CHAMBER PACEMAKER

被引:12
作者
PROVENIER, F [1 ]
VANACKER, R [1 ]
BACKERS, J [1 ]
VANWASSENHOVE, E [1 ]
DEMEYER, V [1 ]
JORDAENS, L [1 ]
机构
[1] ST AUGUSTINUS HOSP,ANTWERP,BELGIUM
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 11期
关键词
ACTIVITY SENSING; DUAL SENSOR; QT INTERVAL; RATE RESPONSIVE PACING;
D O I
10.1111/j.1540-8159.1992.tb02975.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Topaz model 515 (Vitatron B.V.) is a dual sensor rate responsive pacemaker for single chamber stimulation. It can be driven by activity counts (ACT) and QT interval measurements. Inappropriate rate modulation due to one sensor can be corrected by "sensor cross-checking." It was implanted in ten patients (20-86 years) of whom seven had complete heart block and atrial arrhythmias. After implantation T-wave amplitude ranged from 0.9mV-3.5 mV. T-wave sensing ranged from 88%-99% in 9/10 patients at the follow-up of 3 weeks. Eight patients remained in default setting of the activity threshold, after evaluation with a short walking test. An exercise test was performed on all patients. In one test, QT sensing was marginal because of lead implantation in the right ventricular outflow tract. Therefore, this pacing rate was only modulated by ACT sensing. All others were tested with equal contribution of information from both sensors (ACT = QT). In 7/9, rate response was satisfactory. When the treadmill was repeated with ACT in five of these seven patients, rate generally accelerated too fast. In one patient the setting was adjusted to "QT > ACT," because of inappropriate acceleration due to activity sensing, in another it was adjusted to "QT < ACT" because of delayed response to activity. The pacing rate and the ACT during treadmill tests in "QT = ACT" mode were more closely correlated in the first 3 minutes, compared with the last 3 minutes. We feel that rate modulation with this new pacemaker is adequate. Sensor blending and sensor cross-checking are of clinical importance.
引用
收藏
页码:1821 / 1825
页数:5
相关论文
共 6 条
  • [1] HEDMAN A, 1988, PACE, V11, P426
  • [2] INITIAL CLINICAL-EXPERIENCE WITH RATE ADAPTIVE CARDIAC PACING USING 2 SENSORS SIMULTANEOUSLY
    LANDMAN, MAJ
    SENDEN, PJ
    VANROOIJEN, H
    VANHEMEL, NM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1615 - 1622
  • [3] COMBINATION OF SENSORS TO PROVIDE OPTIMAL PACING RATE RESPONSE
    PAUL, V
    GARRATT, C
    CAMM, AJ
    [J]. CLINICAL CARDIOLOGY, 1989, 12 (07) : 400 - 404
  • [4] EXERCISE TESTING IN THE INDICATION AND EVALUATION OF PACEMAKER TREATMENT
    SOWTON, E
    [J]. EUROPEAN HEART JOURNAL, 1987, 8 : 155 - 158
  • [5] Wilkoff BL, 1989, J ELECTROPHYSIOL, V3, P176, DOI DOI 10.1111/J.1540-8167.1989.TB01549.X
  • [6] RATE RESPONSE DURING SUBMAXIMAL EXERCISE - COMPARISON OF 3 DIFFERENT SENSORS
    ZEGELMAN, M
    CIESLINSKI, G
    KREUZER, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1888 - 1895