EVALUATION OF BLOOD-SAMPLING FOR THE MEASUREMENT OF TISSUE-TYPE PLASMINOGEN-ACTIVATOR (T-PA) AND PLASMINOGEN-ACTIVATOR INHIBITOR ACTIVITY (PAI-1)

被引:1
作者
VANDERHEIDE, YT [1 ]
MORRIEN, W [1 ]
VANDENENDE, A [1 ]
机构
[1] UNIV AMSTERDAM,ACAD MED CTR,CTR HEMOSTASIS THROMBOSIS ATHEROSCLEROSIS & INFLA,AMSTERDAM,NETHERLANDS
关键词
ANTICOAGULANTS; TISSUE-TYPE PLASMINOGEN ACTIVATOR; PLASMINOGEN ACTIVATOR INHIBITOR ACTIVITY;
D O I
10.1016/0268-9499(94)90281-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the effect of blood sampling in different anticoagulants on the measurement of t-PA antigen/activity and PAI-1 activity. Blood collection at acid pH (pH 5.9: Stabilyte tubes) yielded higher t-PA activity values (median 0.75 IU/ml, 0.46-1.19) than collection in citrate (median 0.04 IU/ml, 0.01-0.14) or in anticoagulant containing protease inhibitors (median 0.20 IU/ml, 0.02-0.41), No significant effects were seen in the levels of t-PA antigen with the different anticoagulants. For PAI-1 activity determination citrated plasma is preferable to Stabilyte plasma.
引用
收藏
页码:152 / 153
页数:2