CALCULATED VERSUS A-SCAN RESULT FOR AXIAL LENGTH USING DIFFERENT TYPES OF ULTRASOUND PROBE TIP

被引:17
作者
SNEAD, MP
RUBINSTEIN, MP
LEA, SH
HAWORTH, SM
机构
关键词
D O I
10.1038/eye.1990.101
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The axial lengths of 30 pseudophakic eyes were determined by calculation using geometric optics and by A Scan ultrasound using focused fluid filled 'soft' and solid probes. The results from the same ultrasound machine using both types of probe were analysed for their correlation with the calculated axial lengths. The majority of the ultrasound results were shorter than their calculated values but those obtained with the fluid filled 'soft' probe (with consequently less corneal indentation and axial length shortening) were significantly more accurate than those from the solid probe (0.02 > p > 0.01).
引用
收藏
页码:718 / 722
页数:5
相关论文
共 13 条
  • [1] [Anonymous], T OPHTHALMOL SOC UK
  • [2] BINKHORST RD, 1981, OPHTHALMIC SURG LAS, V12, P363
  • [3] Coleman D. J., 1975, BIBL OPHTHALMOL, V83, P246
  • [4] DUANE TD, 1988, CLIN OPHTHALMOL, P4
  • [5] ELDER D, 1970, SYSTEM OPHTHALMOLOGY, V5
  • [6] ELENIUS V, 1963, ACTA OPHTHALMOL, V41, P71
  • [7] JANSSON F, 1962, ACTA OPHTHALMOL, V40, P420
  • [8] JANSSON F, 1963, ACTA OPHTHALMOL, V41, P25
  • [9] PSEUDOPHAKIC ACCOMMODATION - A STUDY OF THE STABILITY OF CAPSULAR BAG SUPPORTED, ONE PIECE, RIGID TRIPOD, OR SOFT FLEXIBLE IMPLANTS
    LEA, SJH
    RUBINSTEIN, MP
    SNEAD, MP
    HAWORTH, SM
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1990, 74 (01) : 22 - 25
  • [10] OLSEN T, 1989, ACTA OPHTHALMOL, V67, P101