EFFECT OF TEMPERATURE AND DENSITY ON THE PERFORMANCE OF MICROPACKED COLUMNS IN SUPERCRITICAL-FLUID CHROMATOGRAPHY

被引:6
作者
IBANEZ, E [1 ]
TABERA, J [1 ]
HERRAIZ, M [1 ]
REGLERO, G [1 ]
机构
[1] CSIC,INST FERMENTAC IND,E-28006 MADRID,SPAIN
关键词
D O I
10.1016/0021-9673(94)89074-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The effects on the efficiency of micropacked columns in supercritical fluid chromatography (SFC) of density at fixed temperature and of temperature at fixed density were investigated. The variation of the retention and resolution achievable for a given pair of solutes with the mobile phase density under different isothermal conditions was also studied. The linear velocity was varied using several flow restrictors of different lengths and inner diameters. The column type investigated was characterized by an inside diameter of <1 mm and the use of large particles (by HPLC standards), resulting in a particle-to-column diameter ratio of 0.1-0.3. Owing to the better permeability derived from both the large particle diameters and the high value of the interparticle porosity, the use of these columns may be an interesting approach in SFC.
引用
收藏
页码:249 / 255
页数:7
相关论文
共 42 条
[1]   PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY WITH 220000 PLATES [J].
BERGER, TA ;
WILSON, WH .
ANALYTICAL CHEMISTRY, 1993, 65 (10) :1451-1455
[2]   EFFECT OF MOBILE PHASE DENSITY GRADIENTS ON EFFICIENCY IN PACKED-COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
BERGER, TA ;
DEYE, JF .
CHROMATOGRAPHIA, 1990, 30 (1-2) :57-60
[3]   EFFECTS OF TEMPERATURE AND DENSITY ON RETENTION IN CAPILLARY SUPERCRITICAL-FLUID CHROMATOGRAPHY [J].
BERGER, TA .
JOURNAL OF CHROMATOGRAPHY, 1989, 478 (02) :311-324
[4]   EFFICIENCY IN PACKED-COLUMN SUPERCRITICAL FLUID CHROMATOGRAPHY USING A MODIFIED MOBILE PHASE [J].
BERGER, TA ;
DEYE, JF .
CHROMATOGRAPHIA, 1991, 31 (11-12) :529-534
[5]   RETENTION IN CAPILLARY SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
CHESTER, TL ;
INNIS, DP .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1985, 8 (09) :561-566
[6]  
CRAMERS CA, 1979, ADV CHROMATOGR, V17, P101
[7]  
Dixon W. J., 1990, BMDP STATISTICAL SOF
[8]   SUPERCRITICAL FLUID CHROMATOGRAPHY WITH SMALL PARTICLE DIAMETER PACKED-COLUMNS [J].
GERE, DR ;
BOARD, R ;
MCMANIGILL, D .
ANALYTICAL CHEMISTRY, 1982, 54 (04) :736-740
[9]   FAILURE OF EDDY DIFFUSION CONCEPT OF GAS CHROMATOGRAPHY [J].
GIDDINGS, JC ;
ROBISON, RA .
ANALYTICAL CHEMISTRY, 1962, 34 (08) :885-&
[10]   EFFECTS OF COLUMN LENGTH, PARTICLE-SIZE, FLOW-RATE, AND PRESSURE PROGRAMMING RATE ON RESOLUTION IN PRESSURE-PROGRAMMED SUPERCRITICAL FLUID CHROMATOGRAPHY [J].
GRAHAM, JA ;
ROGERS, LB .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1980, 18 (02) :75-84