DOSE-RATE FIELD OF THE HERMES-III FLASH X-RAY SOURCE

被引:10
|
作者
SANFORD, TWL
HALBLEIB, JA
MOCK, RC
机构
[1] Sandia National Laboratories, Albuquerque
关键词
D O I
10.1109/23.101189
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Measurements show that the HERMES III flash x-ray source produces a peak dose-rate radiation field of (5 ± 0.5) × 1012 rad(Si)/s over a useful area of 1000 cm2, with a pulse width of ~20 ns. Numerical simulations of the field agree with the measurements and give insight into the dynamics generating the field. The characterization of the radiation field presented here enables test objects to be studied in a well-defined radiation field with unique dose-rate area and volume products. © 1990 IEEE
引用
收藏
页码:1762 / 1768
页数:7
相关论文
共 50 条
  • [1] Design of an X-ray irradiator with FLASH dose-rate capabilities for preclinical research
    Ibanez, P.
    Espinosa-Rodriguez, A.
    Villa-Abaunza, A.
    Sanchez-Parcerisa, D.
    Diaz, N.
    Udias, J. M.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1579 - S1580
  • [2] THE DISTRIBUTION OF DOSE-RATE ACROSS THE FIELD OF AN X-RAY THERAPY TUBE
    FARR, RF
    BRITISH JOURNAL OF RADIOLOGY, 1955, 28 (331): : 364 - 373
  • [3] Design of an X-ray irradiator based on a standard imaging X-ray tube with FLASH dose-rate capabilities for preclinical research
    Espinosa-Rodriguez, A.
    Villa-Abaunza, A.
    Diaz, N.
    Perez-Diaz, M.
    Sanchez-Parcerisa, D.
    Udias, J. M.
    Ibanez, P.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 206
  • [4] SIMULATION OF HIGH DOSE-RATE X-RAY EFFECTS
    HUNTINGTON, JH
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1973, 16 (JUN): : 357 - 357
  • [5] Ultrahigh dose-rate (FLASH) x-ray irradiator for pre-clinical laboratory research
    Rezaee, Mohammad
    Iordachita, Iulian
    Wong, John W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (09):
  • [6] EFFECT OF DOSE-RATE AND FRACTIONATION OF X-RAY DOSE ON ACUTE LETHALITY IN MICE
    BROWN, JAH
    CORP, MJ
    WESTGARTH, DR
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY AND RELATED STUDIES IN PHYSICS CHEMISTRY AND MEDICINE, 1960, 2 (04): : 371 - 381
  • [7] Dosimetric Characterization of the ARIEL 10 MV X-Ray Ultrahigh Dose-Rate (FLASH) Irradiation Platform at TRIUMF
    Esplen, N.
    Egoriti, L.
    Planche, T.
    Hart, A.
    Paley, B.
    Hoehr, C.
    Gottberg, A.
    Bazalova-Carter, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E547 - E547
  • [8] Development of a dose-rate dosimeter for x-ray CT scanner using silicon x-ray diode
    Yamaguchi, Satoshi
    Sato, Eiichi
    Ieko, Yoshiro
    Ariga, Hisanori
    Yoshioka, Kunihiro
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (05):
  • [9] A 5.107-A/KG DOSE-RATE COMPACT X-RAY GENERATOR
    FILATOV, AL
    KOTOV, YA
    MESYATS, GA
    SOKOVNIN, SY
    LASER AND PARTICLE BEAMS, 1989, 7 : 755 - 761
  • [10] High repetition rate portable flash x-ray source
    Viladrosa, R
    Khacef, A
    Cachoncinlle, C
    Pouvesle, JM
    22ND INTERNATIONAL CONGRESS ON HIGH-SPEED PHOTOGRAPHY AND PHOTONICS, 1997, 2869 : 1086 - 1093