FORCES ACTING ON THE MAXILLARY INCISOR TEETH DURING LARYNGOSCOPY USING THE MACINTOSH LARYNGOSCOPE

被引:46
作者
BUCX, MJL
SNIJDERS, CJ
VANGEEL, RTM
ROBERS, C
VANDEGIESSEN, H
ERDMANN, W
STIJNEN, T
机构
[1] ERASMUS UNIV ROTTERDAM,HOSP DIJKZIGT,DEPT BIOMED PHYS & TECHNOL,3015 GD ROTTERDAM,NETHERLANDS
[2] ERASMUS UNIV ROTTERDAM,HOSP DIJKZIGT,DEPT EPIDEMIOL & BIOSTAT,3015 GD ROTTERDAM,NETHERLANDS
[3] ERASMUS UNIV ROTTERDAM,HOSP DIJKZIGT,DEPT CENT INSTRUMENTAT,3015 GD ROTTERDAM,NETHERLANDS
关键词
EQUIPMENT; LARYNGOSCOPE;
D O I
10.1111/j.1365-2044.1994.tb04358.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We determined the forces on the maxillary incisors during routine laryngoscopy in 65 adult patients. The forces were measured by a strain gauge based sensor positioned between the handle and the blade of the laryngoscope. The mean maximal force acting on the maxillary incisors was 49 N. In patients without maxillary incisors, the force acting on the gums was significantly lower at 21 N (p < 0.001). These results suggest that, despite traditional advice to the contrary, a levering movement of the laryngoscope, using the maxillary incisors (ol gums) as a fulcrum, is common practice. Biomechanical analysis revealed that, although levering is not the preferred movement, it is an efficient way of bringing the glottis into view. These results may have implications for future laryngoscope design.
引用
收藏
页码:1064 / 1070
页数:7
相关论文
共 22 条
[1]  
Benumof Jonathan L., 1992, P115
[2]  
BISHOP MJ, 1992, ANESTH ANALG, V74, P411
[3]   MEASUREMENT OF FORCES DURING LARYNGOSCOPY [J].
BUCX, MJL ;
SCHECK, PAE ;
VANGEEL, RTM ;
DENOUDEN, AH ;
NIESING, R .
ANAESTHESIA, 1992, 47 (04) :348-351
[4]   FORCES APPLIED DURING LARYNGOSCOPY IN CHILDREN - ARE VOLATILE ANESTHETICS ESSENTIAL FOR SUXAMETHONIUM INDUCED MUSCLE RIGIDITY [J].
BUCX, MJL ;
VANGEEL, RTM ;
MEURSING, AEE ;
STIJNEN, T ;
SCHECK, PAE .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1994, 38 (05) :448-452
[5]   DENTAL DAMAGE DURING ANESTHESIA AND SURGERY [J].
BURTON, JF ;
BAKER, AB .
ANAESTHESIA AND INTENSIVE CARE, 1987, 15 (03) :262-268
[6]  
CHAFFIN DB, 1984, OCCUPATIONAL BIOMECH, P355
[7]   A MEASURING LARYNGOSCOPE HANDLE - A DEVICE FOR MEASURING THE FORCES APPLIED DURING LARYNGOSCOPY [J].
CHILCOAT, RT ;
ALLEN, FB ;
GERSON, JI ;
GROGONO, AW .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (04) :525-527
[8]   DIFFICULT TRACHEAL INTUBATION IN OBSTETRICS [J].
CORMACK, RS ;
LEHANE, J .
ANAESTHESIA, 1984, 39 (11) :1105-1111
[9]  
DORSCH JA, 1984, UNDERSTANDING ANESTH
[10]  
FISHER TL, 1972, CAN MED ASSOC J, V106, P602