Comparison of central corneal thickness measurements in corneal edema using ultrasound pachymetry, Visante anterior-segment optical coherence tomography, Cirrus optical coherence tomography, and Pentacam Scheimpflug camera tomography

被引:21
作者
Wongchaisuwat, Nida [1 ,2 ]
Metheetrairat, Ankana [1 ]
Chonpimai, Pratuangsri [1 ]
Nujoi, Waree [1 ]
Prabhasawat, Pinnita [1 ]
机构
[1] Mahidol Univ, Siriraj Hosp, Fac Med, Dept Ophthalmol, 2 Prannok Rd, Bangkok 10700, Thailand
[2] Mahidol Univ, Golden Jubilee Med Ctr, Dept Ophthalmol, Salaya, Nakhon Pathom, Thailand
关键词
central corneal thickness; ultrasonography; optical coherence tomography; Scheimpflug camera tomography;
D O I
10.2147/OPTH.S172159
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the central corneal thickness (CCT) measurements in subjects with corneal edema using ultrasound pachymetry, Visante anterior-segment optical coherence tomography (OCT), Cirrus OCT, and Pentacam Scheimpflug camera tomography. Methods: This prospective cross-sectional study included 46 eyes of 33 patients with corneal edema and a CCT exceeding 550 mu m evaluated by ultrasound pachymetry, Visante OCT, Cirrus OCT, and Pentacam. Two observers measured each eye twice. Intraobserver and interobserver reproducibility were determined and agreement among the devices calculated. Results: CCT was measured in 40 eyes of 29 patients. Regardless of the CCT, the measurements obtained using Visante OCT, Cirrus CCT, and ultrasound pachymetry were well correlated. Interobserver and intraobserver reproducibility were high among the three devices. Pentacam overestimated the results compared with the other devices, and ultrasound pachymetry was unmeasurable in six (13%) eyes with very thick and opaque corneas. In eyes with mild corneal edema (CCT 551-650 mu m), measurements from the four devices were comparable. Conclusion: All devices reliably measured the CCT,650 mu m. In eyes with edema exceeding 650 mu m, CCT measurements from the Visante OCT, Cirrus OCT, and ultrasound pachymetry devices showed good reproducibility and were well correlated, while the Pentacam overestimated the values compared to the other devices. Pentacam and ultrasound pachymetry should not be used in eyes with extreme corneal edema and opacity.
引用
收藏
页码:1865 / 1873
页数:9
相关论文
共 21 条
[1]   Comparison and Evaluation of Central Corneal Thickness Using 2 New Noncontact Specular Microscopes and Conventional Pachymetry Devices [J].
Bao, Fangjun ;
Wang, Qinmei ;
Cheng, Shiming ;
Savini, Giacomo ;
Lu, Weicong ;
Feng, Yifan ;
Yu, Ye ;
Huang, Jinhai .
CORNEA, 2014, 33 (06) :576-581
[2]   Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry [J].
Barkana, Y ;
Gerber, Y ;
Elbaz, U ;
Schwartz, S ;
Ken-Dror, G ;
Avni, I ;
Zadok, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1729-1735
[3]   Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter [J].
Bayhan, Hasan Ali ;
Bayhan, Seray Aslan ;
Can, Izzet .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2014, 7 (02) :302-308
[4]  
Chang SW., 2014, TAIWAN J OPHTHALMOL, V4, P163, DOI [DOI 10.1016/J.TJO.2014.08.001, 10.1016/j.tjo.2014.08.001]
[5]   Dependability of Posterior-Segment Spectral Domain Optical Coherence Tomography for Measuring Central Corneal Thickness [J].
Correa-Perez, Maria E. ;
Olmo, Natalia ;
Lopez-Miguel, Alberto ;
Fernandez, Itziar ;
Coco-Martin, Mara B. ;
Maldonado, Miguel J. .
CORNEA, 2014, 33 (11) :1219-1224
[6]   Reproductivity and repeatability of central corneal thickness measurement in keratoconus using the rotating scheimpflug camera and ultrasound pachymetry [J].
De Sanctis, Ugo ;
Missolungi, Alessandro ;
Mutani, Bernardo ;
Richiardi, Lorenzo ;
Grignolo, Federico M. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (05) :712-718
[7]   Graft central thickness measurement by rotating Scheimpflug camera and ultrasound pachymetry after penetrating keratoplasty [J].
de Sanctis, Ugo ;
Missolungi, Alessandro ;
Mutani, Bernardo ;
Grignolo, Federico M. .
OPHTHALMOLOGY, 2007, 114 (08) :1461-1468
[8]   Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry [J].
Grewal, Dilraj S. ;
Brar, Gagandeep S. ;
Grewal, Satinder P. S. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (06) :954-964
[9]   Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia [J].
Ho, Thomas ;
Cheng, Arthur C. K. ;
Rao, Srinivas K. ;
Lau, Silvania ;
Leung, Chris K. S. ;
Lam, Dennis S. C. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (07) :1177-1182
[10]   Central Corneal Thickness Measurements With Fourier-Domain Optical Coherence Tomography Versus Ultrasonic Pachymetry and Rotating Scheimpflug Camera [J].
Ishibazawa, Akihiro ;
Igarashi, Sho ;
Hanada, Kazuomi ;
Nagaoka, Taiji ;
Ishiko, Satoshi ;
Ito, Haruna ;
Yoshida, Akitoshi .
CORNEA, 2011, 30 (06) :615-619