2-COLOR MULTIPARAMETRIC METHOD FOR FLOW CYTOMETRIC DNA ANALYSIS

被引:8
作者
BROWN, RD [1 ]
ZARBO, RJ [1 ]
LINDEN, MD [1 ]
TORRES, FX [1 ]
NAKHLEH, RE [1 ]
SCHULTZ, D [1 ]
MACKOWIAK, PG [1 ]
机构
[1] HENRY FORD HOSP,DIV SURG PATHOL,DETROIT,MI 48202
关键词
COMPENSATION STANDARDS; FLOW CYTOMETRY; HISTOGRAM RESOLUTION; SPECTRAL OVERLAP; SPURIOUS ANEUPLOIDY; 2-COLOR DNA ANALYSIS;
D O I
10.1093/ajcp/101.5.630
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Spectral overlap of green fluorescence signals into the red detector (red-minus-green compensation) is one potential source of variation in two-color flow cytometric DNA analysis. Suboptimal compensation in a two-color propidium iodide (PI)/fluorescein isothiocyanate (FITC) system may be observed if compensation is adjusted using an inappropriate standard, or if changes to fluorescence detector high-voltage settings are made without corresponding readjustment of fluorescence compensation. To quantitate the influence of red-minus-green compensation on the quality of DNA histograms, data from 60 dual PI/cytokeratin (CK)-FITC stained carcinomas were acquired in parallel using two compensation standards: a PI/CK-FITC-stained T24 cell line calibrator overstained to achieve a high-intensity green fluorescence standard (HIGFS) with manually set compensation and automated compensation settings derived from commercial phycoerythrin/low intensity FITC beads (LIGF). Both compensation standards gave similar DNA hyperdiploidy results (DNA index, 1.1-2.8). However, LIGF standard yielded two falsely hypodiploid peaks (DNA index, .7 and .9). Eight left-skewed peaks became DNA diploid and symmetric, respectively, with the HIGFS. Use of HIGFS lowered the coefficient of variation percentage in 95% of cases, the greatest differences (maximum, 3.4%; mean, 1.81%) in tumors of highest intensity CK-FITC. The authors concluded that use of cell-based compensation standards (HIGFS) with intense green signals that mimic clinical tumor samples will avoid spurious aneuploidy and maximize resolution of near-diploid abnormalities.
引用
收藏
页码:630 / 637
页数:8
相关论文
共 7 条
[1]   USE OF QUALITY-CONTROL STANDARDS IN CLINICAL FLOW-CYTOMETRY [J].
AGRAWAL, YP ;
MAHLAMAKI, EK ;
PENTTILA, IM .
ANNALS OF MEDICINE, 1991, 23 (02) :127-133
[2]  
Horan P.K., 1990, FLOW CYTOMETRY SORTI, P397
[3]  
MCCARTHY RC, 1989, ARCH PATHOL LAB MED, V113, P658
[4]  
MCCOY JP, 1994, FLOW CYTOMETRY CLIN, P26
[5]  
ZARBO RJ, 1989, ANAL QUANT CYTOL, V11, P391
[6]   RAPID (ONE-SHOT) STAINING METHOD FOR 2-COLOR MULTIPARAMETRIC DNA FLOW CYTOMETRIC ANALYSIS OF CARCINOMAS USING STAINING FOR CYTOKERATIN AND LEUKOCYTE COMMON ANTIGEN [J].
ZARBO, RJ ;
BROWN, RD ;
LINDEN, MD ;
TORRES, FX ;
NAKHLEH, RE ;
SCHULTZ, DS .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1994, 101 (05) :638-642
[7]  
ZARBO RJ, 1994, FLOW CYTOMETRY CLIN, P425