CHANGES IN OSCILLOMETRIC PULSE AMPLITUDE ENVELOPE WITH CUFF SIZE - IMPLICATIONS FOR BLOOD-PRESSURE MEASUREMENT CRITERIA AND CUFF SIZE SELECTION

被引:16
|
作者
NG, KG
SMALL, CF
机构
[1] Department of Mechanical Engineering, Queen's University, Kingston, Ont. K7L 3N6, McLaughlin Hall
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1993年 / 15卷 / 04期
关键词
BLOOD PRESSURE; NONINVASIVE MEASUREMENT; OSCILLOMETRY; CUFF;
D O I
10.1016/0141-5425(93)90003-H
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillometric blood pressures are derived from the amplitude envelope of oscillometric pulses generated in an occlusive cuff during cuff inflation or deflation; one factor which will affect the characteristics of these pulses is the size of the cuff bladder. Because limiting values are stipulated in recommendations and standards for bladder sizes, there is a wide variety of acceptable cuff sizes for any particular application. An experimental and theoretical study was undertaken to show the dependence of oscillometric blood pressures on bladder size. Actual cuff-arm compliance data were obtained from two subjects for two cuffs of different bladder size. Theoretical analysis was then applied to the data to predict the effects of different bladder sizes on the characteristics of the pulses. The results show that cuff-arm compliance and bladder size interact to affect the pulse amplitude and hence oscillometric blood pressure determination. These results suggest that blood pressures obtained using the oscillometric method may vary depending on cuff size, and in particular that replacement cuffs for oscillometric non-invasive blood pressure monitors should be chosen carefully.
引用
收藏
页码:279 / 282
页数:4
相关论文
共 50 条
  • [21] BLOOD PRESSURE MEASUREMENT: ARM CIRCUMFERENCE AND CUFF SIZE AVAILABILITY
    Veiga, Eugenia Velludo
    Moura Arcuri, Edna Aparecida
    Cloutier, Lyne
    Ferreira Santos, Jair Licio
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2009, 17 (04): : 455 - 461
  • [22] SPHYGMOMANOMETER CUFF SIZE + ACCURACY OF INDIRECT MEASUREMENT OF BLOOD PRESSURE
    KARVONEN, MJ
    TELIVUO, LJ
    JARVINEN, EJ
    AMERICAN JOURNAL OF CARDIOLOGY, 1964, 13 (05): : 688 - &
  • [23] INFLUENCE OF TRANSDUCERS AND OF THE CUFF SIZE ON THE MEASUREMENT OF SYSTOLIC BLOOD-PRESSURE BY ULTRASOUNDS IN NEWBORN-INFANTS
    BALDUCCI, G
    DILECCE, A
    MARGIOTTA, W
    MINERVA PEDIATRICA, 1979, 31 (15) : 1137 - 1140
  • [24] CUFF-LESS BLOOD PRESSURE MEASUREMENT USING THE PULSE TRANSIT TIME - A COMPARISON TO CUFF-BASED OSCILLOMETRIC 24 HOUR BLOOD PRESSURE MEASUREMENT IN CHILDREN
    Hulpke-Wette, M.
    Goehler, A.
    Hofmann, E.
    Kuechler, G.
    JOURNAL OF HYPERTENSION, 2018, 36 : E73 - E73
  • [25] Effect of Respiration on the Characteristic Ratios of Oscillometric Pulse Amplitude Envelope in Blood Pressure Measurement
    Gui, Yihan
    Chen, Fei
    Murray, Alan
    Zheng, Dingchang
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 3646 - 3649
  • [26] BLOOD-PRESSURE MEASUREMENT - GETTING THE RIGHT CUFF
    FEUSSNER, JR
    BLESSINGFEUSSNER, CL
    LINFORS, EW
    NORTH CAROLINA MEDICAL JOURNAL, 1983, 44 (04) : 241 - 241
  • [27] EFFECT OF CUFF LOCATION ON BLOOD-PRESSURE MEASUREMENT
    SHIMIZU, M
    BIORHEOLOGY, 1987, 24 (05) : 515 - 515
  • [28] CUFF PRESSURE OSCILLATIONS IN MEASUREMENT OF RELATIVE BLOOD-PRESSURE
    GEDDES, LA
    NEWBERG, DC
    PSYCHOPHYSIOLOGY, 1977, 14 (02) : 198 - 202
  • [29] NEED FOR AN IMPROVED STANDARD FOR BLOOD-PRESSURE CUFF SIZE - SIZE SHOULD BE RELATED TO DIAMETER OF ARM
    PARK, MK
    KAWABORI, I
    GUNTHEROTH, WG
    CLINICAL PEDIATRICS, 1976, 15 (09) : 784 - 787
  • [30] Blood Pressure Cuff Selection: Does One Size Fit All?
    Jensen, Brock T.
    Walsh, Maureen A.
    Kuhn, Tyler A.
    McFadden, Jessica A.
    Weeter, Taylor M.
    Holmstrup, Michael E.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05): : 909 - 909