LIQUID-CHROMATOGRAPHIC SEPARATION AND STEREOSELECTIVE DETECTION OF L-AMINO AND D-AMINO ACIDS WITH CATALYTIC REACTION DETECTION USING IMMOBILIZED ENZYMES

被引:15
作者
DOMINGUEZ, E [1 ]
MARKOVARGA, G [1 ]
CARLSSON, M [1 ]
GORTON, L [1 ]
机构
[1] UNIV LUND,DEPT ANALYT CHEM,S-22100 LUND,SWEDEN
关键词
D-AMINO ACID OXIDASE; L-AMINO ACID OXIDASE; IMMOBILIZED ENZYME REACTOR; LIQUID CHROMATOGRAPHY; D-AMINO AND L-AMINO ACIDS; SERUM;
D O I
10.1016/0731-7085(90)80127-B
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A post-column LC detection system is described for the stereoselective detection of L- and D-amino acids. The effluent of the LC column passes an immobilized enzyme reactor (IMER) containing either L- or D-amino acid oxidase. The hydrogen peroxide formed in this reactor by the oxidation of the amino acids is then transported to a second IMER containing horse-radish peroxidase. With the addition of 4-aminophenazone and dichlorophenolsulphonyl chloride to the carrier, the hydrogen peroxide is reacted to form a red-coloured complex which is detected in a flow-through photometric cell at 514 nm. Applications to the analysis of amino acids in bovine and human sera are described.
引用
收藏
页码:825 / 830
页数:6
相关论文
共 35 条
[1]   ORTHO-PHTHALALDEHYDE DERIVATIVES OF AMINES FOR HIGH-SPEED LIQUID-CHROMATOGRAPHY ELECTROCHEMISTRY [J].
ALLISON, LA ;
MAYER, GS ;
SHOUP, RE .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :1089-1096
[2]   APPLICATIONS OF IMMOBILIZED BIOCATALYSTS IN CHEMICAL-ANALYSIS [J].
BOWERS, LD .
ANALYTICAL CHEMISTRY, 1986, 58 (04) :A513-&
[3]   GAS-CHROMATOGRAPHIC DETECTION OF D-AMINO ACIDS AS COMMON CONSTITUENTS OF FERMENTED FOODS [J].
BRUCKNER, H ;
HAUSCH, M .
CHROMATOGRAPHIA, 1989, 28 (9-10) :487-492
[4]  
Bycroft B. W., 1988, DICT ANTIBIOTICS REL
[5]   D-AMINO ACIDS IN ANIMALS [J].
CORRIGAN, JJ .
SCIENCE, 1969, 164 (3876) :142-+
[6]   AMINO-ACID ANALYSES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY - AN EVALUATION OF THE USEFULNESS OF PRECOLUMN DNS DERIVATIZATION [J].
DEJONG, C ;
HUGHES, GJ ;
VANWIERINGEN, E ;
WILSON, KJ .
JOURNAL OF CHROMATOGRAPHY, 1982, 241 (02) :345-359
[7]   A FLOW-INJECTION SYSTEM FOR THE AMPEROMETRIC DETERMINATION OF XYLOSE AND XYLULOSE WITH CO-IMMOBILIZED ENZYMES AND A MODIFIED ELECTRODE [J].
DOMINGUEZ, E ;
HAHNHAGERDAL, B ;
MARKOVARGA, G ;
GORTON, L .
ANALYTICA CHIMICA ACTA, 1988, 213 (1-2) :139-150
[8]  
FOX A, 1985, INSTRUMENTAL METHODS, P135
[9]  
FREI RW, 1985, ANAL CHEM, V57, P1529
[10]  
GUBITZ G, 1988, SELECTIVE SAMPLE H A, P154