FLUOROMETRIC-DETERMINATION OF ALKALINE-PHOSPHATASE ACTIVITY IN HUMAN SERUM BY USE OF A FLOW-THROUGH BIOSENSOR

被引:8
作者
SANCHEZCABEZUDO, M
FERNANDEZROMERO, JM
DECASTRO, MDL
机构
[1] UNIV CORDOBA,FAC SCI,DEPT ANALYT CHEM,E-14004 CORDOBA,SPAIN
[2] UNIV COMPLUTENSE MADRID,FAC PHARM,DEPT ANALYT CHEM,INSTRUMENTAL TECH LAB,E-28040 MADRID,SPAIN
关键词
FLOW-THROUGH-BIOSENSOR; FLOW INJECTION; FLUOROMETRY; ALKALINE PHOSPHATASE; ENZYMATIC ACTIVITY; SERUM;
D O I
10.1016/0168-1656(94)90005-1
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A fluorimetric method for the determination of alkaline phosphatase activity based on the use of a flow-through biosensor is reported. The biochemical basis of the method is the hydrolysis of 4-methyl-umbelliferone-phosphate catalyzed by the analyte with fluorimetric monitoring of the 4-methyl-umbelliferone formed (lambda(ex) = 365 nm, lambda(em) = 445 nm). The enhancement of sensitivity achieved when the reaction product is retained on the support packed in the flow-cell makes the method suitable for determination of the analytes in serum samples after 1:50 dilution. The linear range is found to be between 0.1 and 20 U l(-1), with relative standard deviation less than 2.2%. The use of this biosensor was tested by the determination of the analyte in human serum from healthy and sick individuals with excellent recoveries (94-103%).
引用
收藏
页码:143 / 149
页数:7
相关论文
共 27 条
  • [1] HEMOLYSIS AND PLASMA ALKALINE-PHOSPHATASE ACTIVITY
    BALLANTYNE, FC
    BORLAND, W
    BIRRELL, RC
    DAVIDSON, H
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1988, 25 : 192 - 193
  • [2] DENG JT, 1988, CLIN CHIM ACTA, V176, P291
  • [3] FARRERAS P, 1992, MED INTERNA, V1
  • [4] FAULKNER WR, 1986, SELECTED METHODS CLI, V9
  • [5] FLOW-THROUGH OPTICAL BIOSENSOR BASED ON THE PERMANENT IMMOBILIZATION OF AN ENZYME AND TRANSIENT RETENTION OF A REACTION-PRODUCT
    FERNANDEZROMERO, JM
    DECASTRO, MDL
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (21) : 3048 - 3052
  • [6] DETERMINATION OF ENZYMATIC-ACTIVITIES BASED ON AN OPTICAL FLOW-THROUGH P-NITROPHENOL SENSOR
    FERNANDEZROMERO, JM
    DECASTRO, MDL
    VALCARCEL, M
    [J]. ANALYTICAL LETTERS, 1993, 26 (09) : 1847 - 1866
  • [7] FITZPATRICK CP, 1992, CLIN CHEM, V38, P238
  • [8] FITZPATRICK CP, 1992, CLIN CHEM, V38, P247
  • [9] GUILBAULT GG, 1970, ENZYMATIC METHODS AN
  • [10] AMPLIFIED LUMINOMETRIC ASSAYS OF ALKALINE-PHOSPHATASE USING RIBOFLAVIN PHOSPHATES
    HARBRON, S
    EGGELTE, HJ
    BENSON, SM
    RABIN, BR
    [J]. JOURNAL OF BIOLUMINESCENCE AND CHEMILUMINESCENCE, 1991, 6 (04): : 251 - 258