Development of an organ-specific insert phantom generated using a 3D printer for investigations of cardiac computed tomography protocols

被引:39
作者
Abdullah, Kamarul A. [1 ,2 ]
McEntee, Mark F. [1 ]
Reed, Warren [1 ]
Kench, Peter L. [1 ]
机构
[1] Univ Sydney, Discipline Med Radiat Sci, Fac Hlth Sci, Cumberland Campus C42,POB 170, Lidcombe, NSW 1825, Australia
[2] Univ Sultan Zainal Abidin, Fac Hlth Sci, Terengganu, Malaysia
关键词
3D printing; cardiac insert phantom; computed tomography; computer aided design (CAD); rapid prototyping;
D O I
10.1002/jmrs.279
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
IntroductionAn ideal organ-specific insert phantom should be able to simulate the anatomical features with appropriate appearances in the resultant computed tomography (CT) images. This study investigated a 3D printing technology to develop a novel and cost-effective cardiac insert phantom derived from volumetric CT image datasets of anthropomorphic chest phantom. MethodsCardiac insert volumes were segmented from CT image datasets, derived from an anthropomorphic chest phantom of Lungman N-01 (Kyoto Kagaku, Japan). These segmented datasets were converted to a virtual 3D-isosurface of heart-shaped shell, while two other removable inserts were included using computer-aided design (CAD) software program. This newly designed cardiac insert phantom was later printed by using a fused deposition modelling (FDM) process via a Creatbot DM Plus 3D printer. Then, several selected filling materials, such as contrast media, oil, water and jelly, were loaded into designated spaces in the 3D-printed phantom. The 3D-printed cardiac insert phantom was positioned within the anthropomorphic chest phantom and 30 repeated CT acquisitions performed using a multi-detector scanner at 120-kVp tube potential. Attenuation (Hounsfield Unit, HU) values were measured and compared to the image datasets of real-patient and Catphan((R)) 500 phantom. ResultsThe output of the 3D-printed cardiac insert phantom was a solid acrylic plastic material, which was strong, light in weight and cost-effective. HU values of the filling materials were comparable to the image datasets of real-patient and Catphan((R)) 500 phantom. ConclusionsA novel and cost-effective cardiac insert phantom for anthropomorphic chest phantom was developed using volumetric CT image datasets with a 3D printer. Hence, this suggested the printing methodology could be applied to generate other phantoms for CT imaging studies.
引用
收藏
页码:175 / 183
页数:9
相关论文
共 32 条
[1]  
Abdullah KA, 2017, J MED RAD SCI, V64, P93
[2]   Radiation dose and diagnostic image quality associated with iterative reconstruction in coronary CT angiography: A systematic review [J].
Abdullah, Kamarul Amin ;
McEntee, Mark F. ;
Reed, Warren ;
Kench, Peter L. .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2016, 60 (04) :459-468
[3]   Improvements to image quality using hybrid and model-based iterative reconstructions: a phantom study [J].
Aurumskjold, Marie-Louise ;
Ydstrom, Kristina ;
Tingberg, Anders ;
Soderberg, Marcus .
ACTA RADIOLOGICA, 2017, 58 (01) :53-61
[4]  
Austen WG, 1975, CIRCULATION, V51, P5
[5]   Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches [J].
Bae, Kyongtae T. .
RADIOLOGY, 2010, 256 (01) :32-61
[6]   Optimization of the Scan Protocol for the Reduction of Diaphragmatic Motion Artifacts Depicted on CT Angiography: a Phantom Study Simulating Pediatric Patients with Free Breathing [J].
Baek, Jee Hyun ;
Lee, Whal ;
Chang, Kee-Hyun ;
Chung, Jin Wook ;
Park, Jae Hyung .
KOREAN JOURNAL OF RADIOLOGY, 2009, 10 (03) :260-268
[7]   Intravenous contrast material administration at 16-detector row helical CT coronary angiography: Test bolus versus bolus-tracking technique [J].
Cademartiri, F ;
Nieman, K ;
van der Lugt, A ;
Raaijmakers, RH ;
Mollet, N ;
Pattynama, PMT ;
de Feyter, PJ ;
Krestin, GP .
RADIOLOGY, 2004, 233 (03) :817-823
[8]   Low-dose CT coronary angiography using iterative reconstruction with a 256-slice CT scanner [J].
Carrascosa, Patricia ;
Rodriguez-Granillo, Gaston A. ;
Capunay, Carlos ;
Deviggiano, Alejandro .
WORLD JOURNAL OF CARDIOLOGY, 2013, 5 (10) :382-386
[9]   Getting in touch-3D printing in Forensic Imaging [J].
Ebert, Lars Chr. ;
Thali, Michael J. ;
Ross, Steffen .
FORENSIC SCIENCE INTERNATIONAL, 2011, 211 (1-3) :E1-E6
[10]   3D Slicer as an image computing platform for the Quantitative Imaging Network [J].
Fedorov, Andriy ;
Beichel, Reinhard ;
Kalpathy-Cramer, Jayashree ;
Finet, Julien ;
Fillion-Robin, Jean-Christophe ;
Pujol, Sonia ;
Bauer, Christian ;
Jennings, Dominique ;
Fennessy, Fiona ;
Sonka, Milan ;
Buatti, John ;
Aylward, Stephen ;
Miller, James V. ;
Pieper, Steve ;
Kikinis, Ron .
MAGNETIC RESONANCE IMAGING, 2012, 30 (09) :1323-1341