Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

被引:51
|
作者
Gao, Mingwu [1 ]
Olivier, N. Bari [2 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, 428 S Shaw Lane,Rm 2120 Engn Bldg, E Lansing, MI 48824 USA
[2] Michigan State Univ, Dept Small Anim Clin Sci, E Lansing, MI 48824 USA
来源
PHYSIOLOGICAL REPORTS | 2016年 / 4卷 / 10期
关键词
Cuff-less blood pressure; PPG; pulse arrival time; pulse transit time; pulse wave velocity;
D O I
10.14814/phy2.12768
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Pulse transit time (PTT) measured as the time delay between invasive proximal and distal blood pressure (BP) or flow waveforms (invasive PTT [I-PTT]) tightly correlates with BP. PTT estimated as the time delay between non-invasive proximal and distal arterial waveforms could therefore permit cuff-less BP monitoring. A popular noninvasive PTT estimate for this application is the time delay between ECG and photoplethysmography (PPG) waveforms (pulse arrival time [PAT]). Another estimate is the time delay between proximal and distal PPG waveforms (PPG-PTT). PAT and PPG-PTT were assessed as markers of BP over a wide physiologic range using I-PTT as a reference. Waveforms for determining I-PTT, PAT, and PPG-PTT through central arteries were measured from swine during baseline conditions and infusions of various hemodynamic drugs. Diastolic, mean, and systolic BP varied widely in each subject (group average (mean +/- SE) standard deviation between 25 +/- 2 and 36 +/- 2 mmHg). I-PTT correlated well with all BP levels (group average R-2 values between 0.86 +/- 0.03 and 0.91 +/- 0.03). PPG-PTT also correlated well with all BP levels (group average R-2 values between 0.81 +/- 0.03 and 0.85 +/- 0.02), and its R-2 values were not significantly different from those of I-PTT. PAT correlated best with systolic BP (group average R-2 value of 0.70 +/- 0.04), but its R-2 values for all BP levels were significantly lower than those of I-PTT (P < 0.005) and PPG-PTT (P < 0.02). The pre-ejection period component of PAT was responsible for its inferior correlation with BP. In sum, PPG-PTT was not different from I-PTT and superior to the popular PAT as a marker of BP.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Blood pressure monitoring during exercise: Comparison of pulse transit time and volume clamp methods
    Wibmer, Thomas
    Denner, Coy
    Fischer, Christoph
    Schildge, Benedikt
    Ruediger, Stefan
    Kropf-Sanchen, Cornelia
    Rottbauer, Wolfgang
    Schumann, Christian
    BLOOD PRESSURE, 2015, 24 (06) : 353 - 360
  • [42] A novel method for non-invasive blood pressure estimation based on continuous pulse transit time: An observational study
    Shin, Hangsik
    PSYCHOPHYSIOLOGY, 2023, 60 (02)
  • [43] PULSE TRANSIT-TIME - RELATIONSHIP TO BLOOD-PRESSURE AND MYOCARDIAL PERFORMANCE
    OBRIST, PA
    LIGHT, KC
    MCCUBBIN, JA
    HUTCHESON, JS
    HOFFER, JL
    PSYCHOPHYSIOLOGY, 1979, 16 (03) : 292 - 301
  • [44] Using noninvasive adjusted pulse transit time for tracking beat-to-beat systolic blood pressure during ventricular arrhythmia
    Miao, Fen
    Zhou, Bin
    Liu, Zengding
    Wen, Bo
    Li, Ye
    Tang, Min
    HYPERTENSION RESEARCH, 2022, 45 (03) : 424 - 435
  • [45] Using noninvasive adjusted pulse transit time for tracking beat-to-beat systolic blood pressure during ventricular arrhythmia
    Fen Miao
    Bin Zhou
    Zengding Liu
    Bo Wen
    Ye Li
    Min Tang
    Hypertension Research, 2022, 45 : 424 - 435
  • [46] Implantable Reflectance Pulse Transit Time Blood Pressure Sensor with Oximetry Capability
    Fiala, J.
    Gehrke, R.
    Theodor, M.
    Bingger, P.
    Frster, K.
    Heilmann, C.
    Beyersdorf, F.
    Zappe, H.
    Seifert, A.
    BIOPHOTONICS: PHOTONIC SOLUTIONS FOR BETTER HEALTH CARE II, 2010, 7715
  • [47] The acute effects of running on blood pressure estimation using pulse transit time in normotensive subjects
    Wong, Mico Yee-Man
    Pickwell-MacPherson, Emma
    Zhang, Yuan-Ting
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2009, 107 (02) : 169 - 175
  • [48] Case study of ECG signal used as a reference signal in optical pulse transit time measurement of blood flow - The effect of different electrode placements on pulse transit time
    Myllyla, Teemu S.
    Vihriala, Erkki V.
    Korhonen, Vesa O.
    Sorvoja, Hannu S. S.
    SARATOV FALL MEETING 2012: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE XIV; AND LASER PHYSICS AND PHOTONICS XIV, 2013, 8699
  • [49] Continuous Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio
    Ding, Xiao-Rong
    Zhang, Yuan-Ting
    Liu, Jing
    Dai, Wen-Xuan
    Tsang, Hon Ki
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (05) : 964 - 972
  • [50] Methods for reliable estimation of pulse transit time and blood pressure variations using smartphone sensors
    Dias, Alair, Jr.
    Murali, Srinivasan
    Rincon, Francisco
    Atienza, David
    MICROPROCESSORS AND MICROSYSTEMS, 2016, 46 : 84 - 95