Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference

被引:51
|
作者
Gao, Mingwu [1 ]
Olivier, N. Bari [2 ]
Mukkamala, Ramakrishna [1 ]
机构
[1] Michigan State Univ, Dept Elect & Comp Engn, 428 S Shaw Lane,Rm 2120 Engn Bldg, E Lansing, MI 48824 USA
[2] Michigan State Univ, Dept Small Anim Clin Sci, E Lansing, MI 48824 USA
来源
PHYSIOLOGICAL REPORTS | 2016年 / 4卷 / 10期
关键词
Cuff-less blood pressure; PPG; pulse arrival time; pulse transit time; pulse wave velocity;
D O I
10.14814/phy2.12768
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Pulse transit time (PTT) measured as the time delay between invasive proximal and distal blood pressure (BP) or flow waveforms (invasive PTT [I-PTT]) tightly correlates with BP. PTT estimated as the time delay between non-invasive proximal and distal arterial waveforms could therefore permit cuff-less BP monitoring. A popular noninvasive PTT estimate for this application is the time delay between ECG and photoplethysmography (PPG) waveforms (pulse arrival time [PAT]). Another estimate is the time delay between proximal and distal PPG waveforms (PPG-PTT). PAT and PPG-PTT were assessed as markers of BP over a wide physiologic range using I-PTT as a reference. Waveforms for determining I-PTT, PAT, and PPG-PTT through central arteries were measured from swine during baseline conditions and infusions of various hemodynamic drugs. Diastolic, mean, and systolic BP varied widely in each subject (group average (mean +/- SE) standard deviation between 25 +/- 2 and 36 +/- 2 mmHg). I-PTT correlated well with all BP levels (group average R-2 values between 0.86 +/- 0.03 and 0.91 +/- 0.03). PPG-PTT also correlated well with all BP levels (group average R-2 values between 0.81 +/- 0.03 and 0.85 +/- 0.02), and its R-2 values were not significantly different from those of I-PTT. PAT correlated best with systolic BP (group average R-2 value of 0.70 +/- 0.04), but its R-2 values for all BP levels were significantly lower than those of I-PTT (P < 0.005) and PPG-PTT (P < 0.02). The pre-ejection period component of PAT was responsible for its inferior correlation with BP. In sum, PPG-PTT was not different from I-PTT and superior to the popular PAT as a marker of BP.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method
    Gesche, Heiko
    Grosskurth, Detlef
    Kuechler, Gert
    Patzak, Andreas
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (01) : 309 - 315
  • [32] An implantable optical blood pressure sensor based on pulse transit time
    Jens Fiala
    Philipp Bingger
    Dominic Ruh
    Katharina Foerster
    Claudia Heilmann
    Friedhelm Beyersdorf
    Hans Zappe
    Andreas Seifert
    Biomedical Microdevices, 2013, 15 : 73 - 81
  • [33] Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio
    Gholamhosseini, Hamid
    Baig, Mirza
    Rastegar, Solmaz
    Linden, Maria
    PROCEEDINGS OF THE 15TH INTERNATIONAL CONFERENCE ON WEARABLE MICRO AND NANO TECHNOLOGIES FOR PERSONALIZED HEALTH (PHEALTH 2018), 2018, 249 : 77 - 83
  • [34] Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation
    Joonnyong Lee
    JangJay Sohn
    Jonghyun Park
    SeungMan Yang
    Saram Lee
    Hee Chan Kim
    BioMedical Engineering OnLine, 17
  • [35] Smartphone-based Continuous Blood Pressure Measurement Using Pulse Transit Time
    Gholamhosseini, Hamid
    Meintjes, Andries
    Baig, Mirza
    Linden, Maria
    PHEALTH 2016, 2016, 224 : 84 - 89
  • [36] Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation
    Lee, Joonnyong
    Sohn, JangJay
    Park, Jonghyun
    Yang, SeungMan
    Lee, Saram
    Kim, Hee Chan
    BIOMEDICAL ENGINEERING ONLINE, 2018, 17
  • [37] Using Individualized Pulse Transit Time Calibration to Monitor Blood Pressure During Exercise
    Jeong, In Cheol
    Wood, Jeffrey
    Finkelstein, Joseph
    INFORMATICS, MANAGEMENT AND TECHNOLOGY IN HEALTHCARE, 2013, 190 : 39 - 41
  • [38] Contactless video-based photoplethysmography technique comparison investigating pulse transit time estimation of arterial blood pressure
    Shirbani, Fatemeh
    Moriarty, Aidan
    Hui, Nicholas
    Cox, James
    Tan, Isabella
    Avolio, Alberto P.
    Butlin, Mark
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 5650 - 5653
  • [39] Design and Development of Pulse Transit Time Based Cuffless Blood Pressure Monitoring System
    Singh, Harinderjit
    Singh, Mandeep
    2015 INTERNATIONAL CONFERENCE ON ELECTRICAL, ELECTRONICS, SIGNALS, COMMUNICATION AND OPTIMIZATION (EESCO), 2015,
  • [40] Pulse Transit Time Measurement Using Bioimpedance
    Pires, Andrei L. A.
    Souza, Marcio N.
    XXVI BRAZILIAN CONGRESS ON BIOMEDICAL ENGINEERING, CBEB 2018, VOL 1, 2019, 70 (01): : 531 - 535