HOMOGENEOUS AMPEROMETRIC IMMUNOASSAY FOR THEOPHYLLINE IN WHOLE-BLOOD

被引:24
作者
ATHEY, D
MCNEIL, CJ
BAILEY, WR
HAGER, HJ
MULLEN, WH
RUSSELL, LJ
机构
[1] UNIV NEWCASTLE UPON TYNE, SCH MED, DEPT CLIN BIOCHEM, NEWCASTLE UPON TYNE NE2 4HH, ENGLAND
[2] CAMBRIDGE LIFE SCI PLC, ELY CB7 4DT, ENGLAND
关键词
HOMOGENEOUS AMPEROMETRIC IMMUNOASSAY; THEOPHYLLINE; PLATINIZED ACTIVATED CARBON; NADH; EMIT;
D O I
10.1016/0956-5663(93)80026-L
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A homogeneous spectrophotometric EMIT immunoassay kit for the quantitation of theophylline in serum or plasma has been modified to produce a rapid, amperometric immunoassay requiring a 50 mu l whole blood sample. The basis of the detection system for the assay is the electrochemical oxidation of NADH produced by G6PDH-labelled theophylline at a potential of +150 mV vs Ag/AgCl using platinised activated carbon (PACE) electrodes. Comparison of the amperometric whole blood method with the conventional spectrophotometric plasma assay produced a reasonable correlation: y = 0.90x - 1.01, (r = 0.98, n = 12). The advantage of the new method is that simple and robust instrumentation can rapidly determine theophylline in whole blood with no sample pre-treatment or separation steps.
引用
收藏
页码:415 / 419
页数:5
相关论文
共 15 条
[1]   AN ORGANIC CONDUCTOR ELECTRODE FOR THE OXIDATION OF NADH [J].
ALBERY, WJ ;
BARTLETT, PN .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1984, (04) :234-236
[2]   ENZYMATIC DETERMINATION OF GLUCOSE IN A FLOW SYSTEM BY CATALYTIC-OXIDATION OF THE NICOTINAMIDE COENZYME AT A MODIFIED ELECTRODE [J].
APPELQVIST, R ;
MARKOVARGA, G ;
GORTON, L ;
TORSTENSSON, A ;
JOHANSSON, G .
ANALYTICA CHIMICA ACTA, 1985, 169 (MAR) :237-247
[3]   STRATEGIES FOR THE DEVELOPMENT OF AMPEROMETRIC ENZYME ELECTRODES [J].
BARTLETT, PN ;
WHITAKER, RG .
BIOSENSORS, 1988, 3 (06) :359-379
[4]  
BENNETTO HP, 1988, Patent No. 289344
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Cardosi M., 1988, International Industrial Biotechnology, V8, P6
[7]   ELECTROCATALYSIS OF ASCORBATE AND NADH AT A SURFACE MODIFIED GRAPHITE EPOXY ELECTRODE [J].
FALAT, L ;
CHENG, HY .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1983, 157 (02) :393-397
[8]  
HEINEMAN WR, 1987, METHOD BIOCHEM ANAL, V32, P345
[9]  
HOWANITZ PJ, 1984, CLIN DIAGNOSIS MANAG, P346
[10]  
KOYSOOKO R, 1974, CLIN PHARMACOL THER, V15, P454