ELECTROPHORESIS UNDERESTIMATES THE CONCENTRATION OF POLYCLONAL IMMUNOGLOBULINS IN SERUM

被引:7
作者
SCHREIBER, WE [1 ]
CHIANG, E [1 ]
TSE, SSL [1 ]
机构
[1] UNIV BRITISH COLUMBIA,VANCOUVER V6T 1W5,BC,CANADA
关键词
D O I
10.1093/ajcp/97.5.610
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The concentration of gamma-migrating immunoglobulins in blood serum was measured in 200 healthy blood donors by two methods, nephelometry and electrophoresis. Nephelometric values were calculated as IgG + IgM + 1/2 IgA because immunoglobulin G (IgG) and immunoglobulin M (IgM) migrate entirely within the gamma-region and immunoglobulin A (IgA) is split between the gamma and beta-regions. Electrophoretic values were calculated as % gamma X total protein. Gels were scanned with two different densitometers, providing two sets of electrophoretic data. In each case, the correlation between nephelometry and electrophoresis was very good (r = 0.94 and r = 0.96). However, electrophoresis consistently gave lower results than nephelometry (xBAR = 2.3 g/L and 4.8 g/L lower, respectively). The disparity between these methods was greater as the concentration of immunoglobulins increased. It is concluded that (1) electrophoresis gives a relative but not absolute value of immunoglobulin concentration in serum; (2) nephelometric and electrophoretic values cannot be used interchangeably, and (3) reference ranges for electrophoresis must take into account the densitometer used.