FIXATION OF UNSTABLE TROCHANTERIC HIP-FRACTURES - A CADAVER STUDY COMPARING 3 DIFFERENT DEVICES

被引:18
作者
LARSSON, S [1 ]
ELLOY, M [1 ]
HANSSON, LI [1 ]
机构
[1] WRIGHTINGTON HOSP, BIOMECH LAB, WIGAN, ENGLAND
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1988年 / 59卷 / 06期
关键词
D O I
10.3109/17453678809149419
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A four-fragment trochanteric fracture was produced in 24 human cadaver femora. After reduction to anatomic position, the fractures were stabilized by either a Jewett 135.degree. nail-plate, a 135.degree. Nolok sliding screw-plate without a key, or a 140.degree. Hansson pin-plate with an additional trochanteric plate. Using a hip-force simulator, the preparations were subjected to 10,000 loading cycles simulating partial weight bearing, followed by another 10,000 cycles with full weight bearing, while measuring the elastic and permanent fracture displacements. Failure occurred in 3/8 of the Jewett stabilized fractures and in 1/8 each of the NolokTM and Hansson stabilized specimens. Each failure was preceded by an increase in elastic tilt and rotation followed by a permanent rotation after which failure occurred as varus angulation of the fracture. Further, two Jewett nails, one Nolokscrew, and three Hansson pins showed bending. Due to the controlled telescopic action in the Nolok and Hansson dynamic devices, the resistance to fatigue when subjected to repetitive loadings was increased and thereby a better overall results was achieved when compared with the rigid Jewen nail-plate. To maintain the dynamic function, the strength of the dynamic screw-pin must be sufficient to withstand physiologic loadings without bending; otherwise, the telescopic action will be obstructed.
引用
收藏
页码:658 / 663
页数:6
相关论文
共 15 条
[1]  
ELLOY MA, 1977, THESIS U LIVERPOOL L
[2]   AXIAL-COMPRESSION IN FEMORAL-NECK OSTEOTOMIES - A BIOMECHANIC STUDY IN HUMAN CADAVER HIPS [J].
FRANDSEN, PA ;
MADSEN, T .
ACTA ORTHOPAEDICA SCANDINAVICA, 1983, 54 (05) :703-707
[3]   TREATMENT OF INTERTROCHANTERIC FRACTURES OF THE FEMUR - A COMPARISON OF THE RICHARDS SCREW-PLATE WITH THE JEWETT NAIL-PLATE [J].
HEYSEMOORE, GH ;
MACEACHERN, AG ;
EVANS, DCJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1983, 65 (03) :262-267
[4]   SLIDING SCREW IN THE TREATMENT OF TROCHANTERIC AND SUBTROCHANTERIC FRACTURES [J].
HOGH, J .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1982, 14 (02) :141-145
[5]   TREATMENT OF INTERTROCHANTERIC HIP FRACTURES WITH A COMPRESSION HIP SCREW AND A NAIL PLATE [J].
JACOBS, RR ;
ARMSTRONG, HJ ;
WHITAKER, JH ;
PAZELL, J .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1976, 16 (08) :599-603
[6]  
JENSEN JS, 1981, ACTA ORTHOP SCAND, V52, P5
[7]   STABLE FIXATION OF INTERTROCHANTERIC FRACTURES - BIOMECHANICAL EVALUATION [J].
KAUFER, H ;
MATTHEWS, LS ;
SONSTEGARD, D .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1974, A 56 (05) :899-907
[8]   BIOMECHANICAL ANALYSIS OF THE SLIDING CHARACTERISTICS OF COMPRESSION HIP SCREWS [J].
KYLE, RF ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (08) :1308-1314
[9]   FIXATION OF TROCHANTERIC HIP-FRACTURES - A CADAVER STUDY OF STATIC AND DYNAMIC LOADING [J].
LARSSON, S ;
ELLOY, M ;
HANSSON, LI .
ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (04) :365-368
[10]   STABILITY OF OSTEOSYNTHESIS IN TROCHANTERIC FRACTURES - COMPARISON OF 3 FIXATION DEVICES IN CADAVERS [J].
LARSSON, S ;
ELLOY, M ;
HANSSON, LI .
ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (04) :386-390