EXPERIMENTAL SOLID-ANGLE CORRECTION FOR GAMMA-ANGULAR CORRELATIONS

被引:5
作者
VEISSID, VLCP
KENCHIAN, G
DACRUZ, MTF
机构
[1] Instituto de Física, Universidade de São Paulo
关键词
D O I
10.1016/0168-9002(91)90106-Z
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new method is described for performing solid-angle corrections on gamma angular correlation data. The gamma scanning technique was applied to two wrap-around Ge(Li) detectors in order to map their intrinsic photopeak efficiencies as functions of the energy and direction of incidence. A collimated Eu-152 radioactive source of 1 mCi was used in the mapping procedure. Concerning directional correlations, no numerical integration is necessary for each gamma cascade measured, because a different correlation function is proposed for the fitting, where the angle-dependent effects to be corrected have been factorized. Only few energy interpolations are needed, in order to obtain the correction coefficients present in the new function. Tests were performed with the 4 --> 2 --> 0 cascade of Ni-60, and 0' --> 2 --> 0 and 2'--> 2 --> 0 of Ru-100, showing good agreement with previous results.