Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography

被引:41
|
作者
Solomon, Justin [1 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ Hlth Syst, Dept Radiol, Carl E Ravin Adv Imaging Labs, 2424 Erwin Rd,Suite 302, Durham, NC 27705 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Clin Imaging Phys Grp, 2424 Erwin Rd,Suite 302, Durham, NC 27705 USA
[3] Duke Univ, Pratt Sch Engn, Dept Biomed Engn, 2424 Erwin Rd,Suite 302, Durham, NC 27705 USA
[4] Duke Univ, Pratt Sch Engn, Dept Elect & Comp Engn, 2424 Erwin Rd,Suite 302, Durham, NC 27705 USA
关键词
image quality; computed tomography; observer models; detectability;
D O I
10.1117/1.JMI.3.3.035506
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to compare computed tomography (CT) low-contrast detectability from human readers with observer model-based surrogates of image quality. A phantom with a range of low-contrast signals (five contrasts, three sizes) was imaged on a state-of-the-art CT scanner (Siemens' force). Images were reconstructed using filtered back projection and advanced modeled iterative reconstruction and were assessed by 11 readers using a two alternative forced choice method. Concurrently, contrast-to-noise ratio (CNR), area-weighted CNR (CNRA), and observer model-based metrics were estimated, including nonprewhitening (NPW) matched filter, NPW with eye filter (NPWE), NPW with internal noise, NPW with an eye filter and internal noise (NPWEi), channelized Hotelling observer (CHO), and CHO with internal noise (CHOi). The correlation coefficients (Pearson and Spearman), linear discriminator error, E, and magnitude of confidence intervals, vertical bar CI95%vertical bar, were used to determine correlation, proper characterization of the reconstruction algorithms, and model precision, respectively. Pearson (Spearman) correlation was 0.36 (0.33), 0.83 (0.84), 0.84 (0.86), 0.86 (0.88), 0.86 (0.91), 0.88 (0.90), 0.85 (0.89), and 0.87 (0.84), E was 0.25, 0.15, 0.2, 0.25, 0.3, 0.25, 0.4, and 0.45, and vertical bar CI95%vertical bar was 2.84 x 10(-3), 5.29 x 10(-3), 4.91 x 10(-3), 4.55 x 10(-3), 2.16 x 10(-3), 1.24 x 10(-3), 4.58 x 10-2, and 7.95 x 10-2 for CNR, CNRA, NPW, NPWE, NPWi, NPWEi, CHO, and CHOi, respectively. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Impact of Model-Based Iterative Reconstruction on the Correlation between Computed Tomography Quantification of a Low Lung Attenuation Area and Airway Measurements and Pulmonary Function Test Results in Normal Subjects
    Kim, Da Jung
    Kim, Cherry
    Shin, Chol
    Lee, Seung Ku
    Ko, Chang Sub
    Lee, Ki Yeol
    KOREAN JOURNAL OF RADIOLOGY, 2018, 19 (06) : 1187 - 1195
  • [22] Anthropomorphic model observer performance in three-dimensional detection task for low-contrast computed tomography
    Ba, Alexandre
    Eckstein, Miguel P.
    Racine, Damien
    Ott, Julien G.
    Verdun, Francis
    Kobbe-Schmidt, Sabine
    Bochud, Francois O.
    JOURNAL OF MEDICAL IMAGING, 2016, 3 (01)
  • [23] CNN as model observer in a liver lesion detection task for x-ray computed tomography: A phantom study
    Kopp, Felix K.
    Catalano, Marco
    Pfeiffer, Daniela
    Fingerle, Alexander A.
    Rummeny, Ernst J.
    Noel, Peter B.
    MEDICAL PHYSICS, 2018, 45 (10) : 4439 - 4447
  • [24] Learning effects in visual grading assessment of model-based reconstruction algorithms in abdominal Computed Tomography
    Kataria, Bharti
    Oman, Jenny
    Sandborg, Michael
    Smedby, Orjan
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2023, 10
  • [25] A qualitative study of improving megavoltage computed tomography image quality and maintaining dose accuracy using cycleGAN-based image synthesis
    Lv, Tie
    Xie, Chuanbin
    Zhang, Yihang
    Liu, Yaoying
    Zhang, Gaolong
    Qu, Baolin
    Zhao, Wei
    Xu, Shouping
    MEDICAL PHYSICS, 2024, 51 (01) : 394 - 406
  • [26] Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients
    C Burgstahler
    T Beck
    A Kuettner
    A Reimann
    A F Kopp
    M Heuschmid
    C D Claussen
    S Schroeder
    International Journal of Obesity, 2006, 30 : 569 - 573
  • [27] Image quality and diagnostic accuracy of 16-slice multidetector computed tomography for the detection of coronary artery disease in obese patients
    Burgstahler, C
    Beck, T
    Kuettner, A
    Reimann, A
    Kopp, AF
    Heuschmid, M
    Claussen, CD
    Schroeder, S
    INTERNATIONAL JOURNAL OF OBESITY, 2006, 30 (03) : 569 - 573
  • [28] Observer- and radiomics model-based computed tomography classification of suppurative versus tuberculous lymphadenitis complicated with nodal necrosis of the neck in children
    Rui Zhang
    Ye Xu
    Sijie Gao
    Yang Jing
    Wei Li
    Pediatric Radiology, 2023, 53 : 2586 - 2596
  • [29] Observer- and radiomics model-based computed tomography classification of suppurative versus tuberculous lymphadenitis complicated with nodal necrosis of the neck in children
    Zhang, Rui
    Xu, Ye
    Gao, Sijie
    Jing, Yang
    Li, Wei
    PEDIATRIC RADIOLOGY, 2023, 53 (13) : 2586 - 2596
  • [30] Medical Image Analysis and Correlation Between Ankle Fracture Classification and Ankle Computed Tomography
    Teng, Yugang
    Zhang, Yuanzhen
    Wang, Zhenyu
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2020, 10 (12) : 2935 - 2939