BIOCOMPATIBILITY OF HEPARIN-COATED CIRCUITS USED IN CARDIOPULMONARY BYPASS

被引:22
作者
PEKNA, M
BOROWIEC, J
FAGERHOL, MK
VENGE, P
THELIN, S
机构
[1] UNIV UPPSALA HOSP, DEPT THORAC & CARDIOVASC SURG, S-75185 UPPSALA, SWEDEN
[2] UNIV UPPSALA HOSP, DEPT CLIN IMMUNOL & TRANSFUS MED, S-75185 UPPSALA, SWEDEN
[3] UNIV UPPSALA HOSP, DEPT CLIN CHEM, S-75185 UPPSALA, SWEDEN
[4] ULLEVAL UNIV HOSP, BLOOD BANK, OSLO, NORWAY
[5] ULLEVAL UNIV HOSP, DEPT IMMUNOL, OSLO, NORWAY
来源
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY | 1994年 / 28卷 / 01期
关键词
CARDIOPULMONARY BYPASS; HEPARIN COATING; LEUKOCYTE ACTIVATION; COMPLEMENT ACTIVATION;
D O I
10.3109/14017439409098703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The combined effect of heparin coating of cardiopulmonary bypass (CPB) circuits and reduced dose of systemic heparin on activation of the complement system and blood leukocytes was investigated in 19 patients undergoing coronary bypass surgery and randomly allocated to two groups. A heparin-coated CPB circuit together with a 50% reduction of the standard heparin dose were used for ten patients (HC group), and a standard CPB circuit with a standard heparin dose (300 IU/kg) for nine (C group). Significant rise in the levels of neutrophil-derived myeloperoxidase, lactoferrin and calprotectin were observed during CPB in both groups, but the total accumulated levels were significantly lower in the HC than in the C group (p < 0.05). Complement activation, assessed from levels of C3a and terminal complement complexes was similar in both groups. The lower levels of myeloperoxidase, lactoferrin and calprotectin during CPB in the HC group indicate that surface modification with endpoint attached heparin enhances the biocompatibility of CPB.
引用
收藏
页码:5 / 11
页数:7
相关论文
共 31 条
  • [1] NEUTROPHIL LYSOSOMAL-ENZYME RELEASE AND COMPLEMENT ACTIVATION DURING CARDIOPULMONARY BYPASS
    ANTONSEN, S
    BRANDSLUND, I
    CLEMENSEN, S
    SOFELDT, S
    MADSEN, T
    ALSTRUP, P
    [J]. SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1987, 21 (01): : 47 - 52
  • [2] BERNTZEN HB, 1989, J RHEUMATOL, V16, P1416
  • [3] DECREASED BLOOD-LOSS AFTER CARDIOPULMONARY BYPASS USING HEPARIN-COATED CIRCUIT AND 50-PERCENT REDUCTION OF HEPARIN DOSE
    BOROWIEC, J
    THELIN, S
    BAGGE, L
    HULTMAN, J
    HANSSON, HE
    [J]. SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1992, 26 (03): : 177 - 185
  • [4] BOROWIEC J, 1992, J THORAC CARDIOV SUR, V104, P642
  • [5] HEPARIN COATING REDUCES BLOOD-CELL ADHESION TO ARTERIAL FILTERS DURING CORONARY-BYPASS - A CLINICAL-STUDY
    BOROWIEC, JW
    BYLOCK, A
    VANDERLINDEN, J
    THELIN, S
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (06) : 1540 - 1545
  • [6] INFLAMMATORY RESPONSE TO CARDIOPULMONARY BYPASS
    BUTLER, J
    ROCKER, GM
    WESTABY, S
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (02) : 552 - 559
  • [7] PLATELET AND NEUTROPHIL ACTIVATION IN CARDIOPULMONARY BYPASS
    COLMAN, RW
    [J]. ANNALS OF THORACIC SURGERY, 1990, 49 (01) : 32 - 34
  • [8] COLTON T, 1974, STAT MED, P207
  • [9] DEPALMA L, 1991, J THORAC CARDIOV SUR, V101, P240
  • [10] Edens E, 1993, COMPLEMENT PROFILES, V1, P96