CALORIMETRIC PARTICLE DETECTORS WITH SUPERCONDUCTING ABSORBER MATERIALS

被引:8
|
作者
FORSTER, G
FERGER, P
VONFEILITZSCH, F
MOSSBAUER, RL
机构
[1] Technische Universität München, Physik-Department E15
关键词
D O I
10.1016/0168-9002(93)91053-P
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We use massive superconducting absorbers made of molybdenum and vanadium as low temperature calorimetric particle detectors. The high resolution of our thermometry system, consisting of a superconducting phase transition thermometer monitored with a dc-SQUID, enables us to detect alpha and gamma particles with large single crystals. Heat pulses generated by passing current through a metal film on the surface of the crystal are used to study the response of the calorimeters. With a 35 g molybdenum single crystal we obtain an energy resolution of 10% FWHM on 5.8 MeV alpha particles at an operating temperature of 120 mK. The observed temperature rise of 1.2 muK is a factor of 8 less than expected from the calculated heat capacity. Using a 15 g vanadium single crystal, the energy resolution on 5.8 MeV a particles is 1.2% FWHM. In this case, the pulse height of 6.1 muK is a factor of 8 smaller than expected from the calculated heat capacity. A possible reason for the large deviations from the expected heat capacity is the presence of hydrogen dissolved in vanadium. Impurities and defects on the surface of the crystals probably contribute also to the heat capacity. The pulse shapes of all observed events are very similar, whether they are created by absorption of alpha or gamma radiation or by an electrically created heat pulse, and consists of two exponential decaying parts. A comparison of the pulse heights of heater and radiation pulses leads us to the conclusion that all the energy deposited in the absorber crystal is converted into phonons.
引用
收藏
页码:491 / 495
页数:5
相关论文
共 50 条
  • [31] MODEL FOR CRYOGENIC PARTICLE DETECTORS WITH SUPERCONDUCTING PHASE-TRANSITION THERMOMETERS
    PROBST, F
    FRANK, M
    COOPER, S
    COLLING, P
    DUMMER, D
    FERGER, P
    FORSTER, G
    NUCCIOTTI, A
    SEIDEL, W
    STODOLSKY, L
    JOURNAL OF LOW TEMPERATURE PHYSICS, 1995, 100 (1-2) : 69 - 104
  • [32] PHONON-MEDIATED PARTICLE DETECTORS - PHYSICS AND MATERIALS
    GIULIANI, A
    SANGUINETTI, S
    MATERIALS SCIENCE & ENGINEERING R-REPORTS, 1993, 11 (01) : 1 - 52
  • [34] CHARACTERIZATION OF THIN-FILM SUPERCONDUCTING DOT ARRAYS FOR CRYOGENIC PARTICLE DETECTORS
    LEONI, R
    CASTELLANO, MG
    SPAGNOLO, GS
    CARELLI, P
    CIRILLO, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (02) : 258 - 262
  • [35] SUPERCONDUCTING TUNNEL JUNCTION DETECTORS - THE QUASI-PARTICLE DIFFUSION IN THE NONLINEAR REGIME
    ESPOSITO, E
    IVLEV, B
    NAPPI, C
    PEPE, G
    DIUCCIO, US
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 355 (2-3) : 515 - 520
  • [36] CHARGE-TO-PHOTON TRANSDUCERS FOR PARTICLE DETECTORS FOR THE SUPERCONDUCTING SUPER-COLLIDER
    BECKER, RA
    KINCAID, BE
    APPLIED OPTICS, 1993, 32 (33): : 6690 - 6697
  • [37] CURRENT-BIASED SUPERCONDUCTING STRIPS AS POSITION-SENSITIVE PARTICLE DETECTORS
    SCHERSCHEL, MA
    HAGEN, CW
    JAGGI, A
    MAIER, T
    ZEHNDER, A
    ZHAO, SP
    OTT, HR
    JOURNAL OF APPLIED PHYSICS, 1995, 77 (06) : 2231 - 2243
  • [38] Mew scintillating materials based on liquid crystals for particle detectors
    Barnik, MI
    Vasil'chenko, VG
    Golovkin, SV
    Medvedkov, AM
    Solov'ev, AS
    Yudin, SG
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1999, 42 (06) : 771 - 777
  • [39] Computer simulations for the design of calorimetric detectors
    Oberstedt, A
    Holmstedt, N
    Tovesson, F
    Belmans, W
    De Moor, P
    Severijns, N
    Vanneste, L
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) : 265 - 267
  • [40] Calorimetric detectors for high mass ions
    University of Oxford, Dept. Phys., Particle Nucl. Phys., Keble Road, Oxford OX1 3RH, United Kingdom
    RAPID COMMUN. MASS SPECTROM., 2 (171-174):