AN INVIVO EVALUATION OF ENDEX ELECTRONIC APEX LOCATOR

被引:55
作者
FRANK, AL [1 ]
TORABINEJAD, M [1 ]
机构
[1] LOMA LINDA UNIV,GRAD PROGRAM,LOMA LINDA,CA 92350
关键词
D O I
10.1016/S0099-2399(06)80683-7
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Because of technical problems and hazards of radiation, dentists have desired alternative ways to determine the working length during root canal therapy. Attention has turned to electronic devices for measuring root canal length. Several such devices are now available. They show promise but their accuracy is hampered because moisture in the root canal can result in inaccurate responses. Recently, a new device, Endex, has been introduced, claiming ease of operation and accuracy in moist environments. Ten operators used the Endex to measure 185 root canal lengths (in 99 teeth). Endex measurements were compared with radiographic measurements. The Endex located the apical constriction accurately within +/-0.5 mm in 89.64% of moist canals.
引用
收藏
页码:177 / 179
页数:3
相关论文
共 11 条
  • [1] CUSTER LW, 1916, J NATL DENT ASS, V5, P815
  • [2] FOUAD A, 1989, J ENDODONT, V12, P573
  • [3] HASEGAWA K, 1986, Journal of Nihon University School of Dentistry, V28, P117
  • [4] HULSMANN M, 1989, ENDOD DENT TRAUMATOL, V5, P238
  • [5] A CLINICAL-EVALUATION OF THE ENDOCATER - AN ELECTRONIC APEX LOCATOR
    KELLER, ME
    BROWN, CE
    NEWTON, CW
    [J]. JOURNAL OF ENDODONTICS, 1991, 17 (06) : 271 - 274
  • [6] AN EVALUATION OF THE APEX LOCATOR ENDOCATER
    MCDONALD, NJ
    HOVLAND, EJ
    [J]. JOURNAL OF ENDODONTICS, 1990, 16 (01) : 5 - 8
  • [7] NAHMIS Y, 1987, J ENDODONT, V5, P209
  • [8] ONEILL LJ, 1974, ORAL SURG ORAL MED O, V38, P369
  • [9] SAITO T, 1990, DENT JAPAN, V3, P1
  • [10] NEW METHOD FOR MEASURING LENGTH OF ROOT CANAL
    SUNADA, I
    [J]. JOURNAL OF DENTAL RESEARCH, 1962, 41 (02) : 375 - &