PRESSURIZED AEROSOL VERSUS JET AEROSOL DELIVERY TO MECHANICALLY VENTILATED PATIENTS - COMPARISON OF DOSE TO THE LUNGS

被引:152
作者
FULLER, HD [1 ]
DOLOVICH, MB [1 ]
POSMITUCK, G [1 ]
PACK, WW [1 ]
NEWHOUSE, MT [1 ]
机构
[1] MCMASTER UNIV,ST JOSEPHS HOSP,DEPT NUCL MED,HAMILTON,ONTARIO,CANADA
来源
AMERICAN REVIEW OF RESPIRATORY DISEASE | 1990年 / 141卷 / 02期
关键词
D O I
10.1164/ajrccm/141.2.440
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The purpose of this study was to compare deposition of aerosol to the lung from a metered-dose inhaler (MDI) and aerosol holding chamber and from a jet nebulizer in ventilator-dependent patients. Twenty-one patients were entered into the study, all receiving assisted ventilation and inhaled bronchodilators because of airflow limitation. The average age was 68 yr; there were 10 men and 11 women. The patients were randomized to receive either 4 puffs (800 μg) of radiolabeled fenoterol by MDI of 1.75 ml (1,750 μg) of radiolabeled fenoterol solution by nebulizer. Imaging of lung fields was made by a portable scintillation camera at 5-min intervals during the study. Results showed that 20 patients completed the study, 9 receiving fenoterol by MDI, and 11 by jet nebulizer. Four were exluded from analysis because of previous pneumonectomy, two from each group. Lung deposition measured as a percent of given dose from either system was 5.65 ± 1.09 (mean ± SEM) for MDI plus extension chamber and 1.22 ± 0.35 for jet nebulizer (p < 0.001). Therefore, this trial shows significantly greater efficiency of aerosol deposition to the lung in ventilator-dependent patients when using an MDI plus aerosol holding chamber than when using a jet nebulizer.
引用
收藏
页码:440 / 444
页数:5
相关论文
共 14 条
[1]   DESIGN AND CHARACTERISTICS OF A PORTABLE BREATH ACTUATED, PARTICLE-SIZE SELECTIVE MEDICAL AEROSOL INHALER [J].
CORR, D ;
DOLOVICH, M ;
MCCORMACK, D ;
RUFFIN, R ;
OBMINSKI, G ;
NEWHOUSE, M .
JOURNAL OF AEROSOL SCIENCE, 1982, 13 (01) :1-7
[2]   CLINICAL-EVALUATION OF A SIMPLE DEMAND INHALATION MDI AEROSOL DELIVERY DEVICE [J].
DOLOVICH, M ;
RUFFIN, R ;
CORR, D ;
NEWHOUSE, MT .
CHEST, 1983, 84 (01) :36-41
[3]  
FLAVIN M, 1986, PEDIATR PULM, V2, P36
[4]  
FRASER I, 1981, AM REV RESPIR DIS, V123, P107
[5]  
HUGHES J M, 1987, Respiratory Care, V32, P1131
[6]   COMPARISON OF DOMICILIARY NEBULIZED SALBUTAMOL AND SALBUTAMOL FROM A METERED-DOSE INHALER IN STABLE CHRONIC AIR-FLOW LIMITATION [J].
JENKINS, SC ;
HEATON, RW ;
FULTON, TJ ;
MOXHAM, J .
CHEST, 1987, 91 (06) :804-807
[7]  
KOHLER D, 1984, RESPIRATION S1, V46, P99
[8]  
LEGARE M, 1988, American Review of Respiratory Disease, V137, P60
[9]   AEROSOL DELIVERY IN INTUBATED, MECHANICALLY VENTILATED PATIENTS [J].
MACINTYRE, NR ;
SILVER, RM ;
MILLER, CW ;
SCHULER, F ;
COLEMAN, RE .
CRITICAL CARE MEDICINE, 1985, 13 (02) :81-84
[10]  
MADSEN EB, 1982, EUR J CLIN PHARMACOL, V23, P27