LINEAR POSITION-SENSITIVE X-RAY-DETECTOR INCORPORATING A SELF-SCANNING PHOTO-DIODE ARRAY

被引:21
作者
GAMBLE, RC [1 ]
BALDESCHWIELER, JD [1 ]
GIFFIN, CE [1 ]
机构
[1] CALTECH,JET PROP LAB,PASADENA,CA 91103
关键词
D O I
10.1063/1.1135730
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A linear position-sensitive x-ray detector for x-ray spectroscopy and diffraction applications has been tested which can provide excellent spatial resolution, wide dynamic range and good sensitivity. The heart of the system is a self-scanning, photosensitive silicon diode array. It is interfaced via fiber optics to a thin layer of ZnS which fluoresces visible light upon absorption of x-radiation. The conversion to visible light and optical coupling provide several-fold gain in the efficiency of detection as compared to the direct detection of x-ray by the diode array. Equally important is that the array is protected from irreversible damage by high energy radiation, a limitation which previously hindered this application of silicon diode technology.
引用
收藏
页码:1416 / 1420
页数:5
相关论文
共 18 条
[1]  
CHESTER AM, 1969, BELL SYSTEM TECH FEB, P345
[2]   CHARACTERISTICS OF PHOTODIODE ARRAYS FOR SPECTROCHEMICAL MEASUREMENTS [J].
HORLICK, G .
APPLIED SPECTROSCOPY, 1976, 30 (02) :113-123
[3]   FIBER OPTICS .7. IMAGE TRANSFER FROM LAMBERTIAN EMITTERS [J].
KAPANY, NS ;
CAPELLARO, DF .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1961, 51 (01) :23-&
[4]   FIBER OPTICS .8. FOCON [J].
KAPANY, NS .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1961, 51 (01) :32-&
[5]   ACTIVE-RECORDING X-RAY CRYSTAL SPECTROMETER FOR LASER-INDUCED PLASMAS [J].
KOPPEL, LN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1976, 47 (09) :1109-1112
[6]   ABSOLUTE X-RAY EFFICIENCIES OF SOME PHOSPHORS [J].
POORTER, JAD ;
BRIL, A .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1975, 122 (08) :1086-1088
[7]   HIGH SENSITIVITY IMAGE INTENSIFIER TV DETECTOR FOR X-RAY-DIFFRACTION STUDIES [J].
REYNOLDS, GT ;
MILCH, JR ;
GRUNER, SM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1978, 49 (09) :1241-1249
[8]   PHOTOELECTRON COUNTING WITH AN IMAGE INTENSIFIER TUBE AND A SELF-SCANNED PHOTODIODE ARRAY [J].
SANDEL, BR ;
BROADFOOT, AL .
APPLIED OPTICS, 1976, 15 (12) :3111-3114
[9]   RECENT DEVELOPMENTS IN X-RAY AND NEUTRON SMALL-ANGLE SCATTERING INSTRUMENTATION AND DATA-ANALYSIS [J].
SCHELTEN, J ;
HENDRICKS, RW .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1978, 11 (OCT) :297-324
[10]  
STORM E, 1967, LA3753 LOS AL SCI LA