CARDIAC-OUTPUT MEASUREMENT BY THE THERMODILUTION METHOD - AN IN-VITRO TEST OF ACCURACY OF 3 COMMERCIALLY AVAILABLE AUTOMATIC CARDIAC-OUTPUT COMPUTERS

被引:9
作者
RUBINI, A [1 ]
DELMONTE, D [1 ]
CATENA, V [1 ]
ATTAR, I [1 ]
CESARO, M [1 ]
SORANZO, D [1 ]
RATTAZZI, G [1 ]
ALATI, GL [1 ]
机构
[1] GEN HOSP P COSMA,ULSS 20,ICU,PADUA,ITALY
关键词
CARDIAC OUTPUT; THERMODILUTION; MONITORING SYSTEMS; TEST IN VITRO;
D O I
10.1007/BF01726539
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To describe the accuracy and the reproducibility of the thermodilution flow measurements obtained using 3 commercially available cardiac output computers commonly used in intensive care units. Design: An experimental ''in vitro'' study. Twelve different values of control flow (Qctr) were measured (Qmsr) using 3 different cardiac output computers (Abbott Critical Care System, Oximetrix 3 SvO(2)/CO Computer, Baxter Oximeter/Cardiac Output Computer SAT-1(TM); American Edwards Laboratories, 9520A Cardiac Output Computer). Standard equipment and techniques were employed, taking account of the specific weight and heat of warm water relative to blood. In addition, separate sets of measurements were performed in order to investigate the effect on Qmsr of some variables which may influence the ''indicator'' loss (time for injection, depth of immersion of the catheter, temperature of the injected fluid). Setting: Our laboratory, inside the intensive care unit. Measurements and results: The analysis of the linear regression of Qmsr versus Qctr (r values between 0.992 and 0.984; residual standard deviation values comprised between 0.24 and 0.49 1/min; intercepts and slopes not significantly different from identity line), the values of the percentage errors (PE = [Qctr- Qmsr] 100/Qctr; PE mean values 7.9, 5.0 and 13.1), and those of the coefficients of variability (CV = standard deviation mean value, %; CV mean values 5.4, 5.8 and 4.6), show a good level of accuracy and reproducibility of the measurements. Our data confirm previously reported results. Furthermore, the cumulative effect of variables capable of influencing the ''indicator'' loss, even if corrected according to the ''calculation constant'' the manufacturers provide, was found to result in statistically significant changes of Qmsr. Conclusion: The accuracy and reproducibility of the automatic cardiac computers tested is sufficient for practical clinical purpose. It may also depend on the modality of injection of the cooling bolus, which may significantly influence the effective ''indicator'' losses.
引用
收藏
页码:154 / 158
页数:5
相关论文
共 10 条
  • [1] HAND HELD THERMODILUTION CARDIAC-OUTPUT INJECTOR
    DIZON, CT
    GEZARI, WA
    BARASH, PG
    CRITTENDEN, JF
    [J]. CRITICAL CARE MEDICINE, 1977, 5 (04) : 210 - 212
  • [2] MEASUREMENT OF CARDIAC OUTPUT IN ANAESTHETIZED ANIMALS BY A THERMO-DILUTION METHOD
    FEGLER, G
    [J]. QUARTERLY JOURNAL OF EXPERIMENTAL PHYSIOLOGY AND COGNATE MEDICAL SCIENCES, 1954, 39 (03): : 153 - 164
  • [3] FORRESTER JS, 1972, AM HEART J, V3, P306
  • [4] NEW TECHNIQUE FOR MEASUREMENT OF CARDIAC OUTPUT BY THERMODILUTION IN MAN
    GANZ, W
    DONOSO, R
    MARCUS, HS
    FORRESTER, JS
    SWAN, HJC
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1971, 27 (04) : 392 - +
  • [5] COMPARISON OF 2 CLOSED SYSTEMS FOR THERMODILUTION CARDIAC OUTPUTS
    PLACHETKA, JR
    LARSON, DF
    SALOMON, NW
    COPELAND, JG
    [J]. CRITICAL CARE MEDICINE, 1981, 9 (06) : 487 - 489
  • [6] POWNER DJ, 1978, MED INSTRUM, V12, P122
  • [7] SAADJIAN A, 1976, MED PROG TECHNOL, V3, P161
  • [8] MEASUREMENT OF CARDIAC-OUTPUT BY THERMODILUTION
    WEISEL, RD
    BERGER, RL
    HECHTMAN, HB
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1975, 292 (13) : 682 - 684
  • [9] EVALUATION OF THERMAL DILUTION FOR OBTAINING SERIAL MEASUREMENTS OF CARDIAC-OUTPUT
    WILSON, EM
    UPDIKE, OL
    RANIERI, AJ
    DAMMANN, JF
    [J]. MEDICAL & BIOLOGICAL ENGINEERING, 1972, 10 (02): : 179 - &
  • [10] 1929, INT CRITICAL TABLES