DETERMINATION OF URANIUM IN ORGANIC SOLUTIONS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY
被引:0
作者:
ILIC, Z
论文数: 0引用数: 0
h-index: 0
ILIC, Z
MILONJIC, S
论文数: 0引用数: 0
h-index: 0
MILONJIC, S
COMOR, J
论文数: 0引用数: 0
h-index: 0
COMOR, J
机构:
来源:
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES
|
1992年
/
158卷
/
01期
关键词:
D O I:
10.1007/BF02034769
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The determination of uranium in organic solutions by inductively coupled plasma atomic emission spectrometry is reported. The conditions for achieving plasma stability were determined as well as the optimum conditions (RF power, observation height and argon flow) of a spectrometer for analysis of U(II) 367.007 nm in three organic solvents, xylene, kerosine and benzene. The effects of RF power and observation height on signal-to-background ratio and the detection limits are considered. Spectral interferences due to CN band are discussed.