SPECKLE REDUCTION IN PULSE ECHO ULTRASONIC-IMAGING USING A 2-DIMENSIONAL RECEIVING ARRAY

被引:6
作者
GIESEY, JJ
CARSON, PL
FITTING, DW
MEYER, CR
机构
[1] UNIV MICHIGAN HOSP,DEPT RADIOL,ANN ARBOR,MI 48109
[2] NATL INST STAND & TECHNOL,DIV MAT RELIABIL,BOULDER,CO 80303
关键词
D O I
10.1109/58.139111
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An experimental pulse-echo imager was developed for the purpose of reducing speckle in ultrasonic images. The system utilized a 64 element spherically focused segmented annuli array receiver with a common transmitter. Compounded images were formed using subapertures of varying size, shape and overlap, and the speckle and resolution characteristics of the final images observed. A point-like scatterer was imaged to determine the resolution, point spread function and sensitivity of the system along with a new measure called the resolution cell size. The response of the system was also simulated for comparisons. Images were made of a uniformly scattering material to determine the speckle signal-to-noise ratio, the speckle spot size and a measure of image quality, the normalized contrast-to-speckle ratio. Lateral resolution and resolution cell sizes only gradually increased with a decrease in subaperture size and system sensitivity was not greatly diminished. Incoherent summation of signals from small groups of elements, as compared to coherent summation, decreased the speckle noise by a factor of four while maintaining enough resolution to improve the image quality as measured by the CSR/d by a factor of almost two. The subaperture shape and orientation of overlap were also found to be significant.
引用
收藏
页码:167 / 173
页数:7
相关论文
共 19 条
[11]   THE IMPROVEMENT AND QUANTITATIVE ASSESSMENT OF B-MODE IMAGES PRODUCED BY AN ANNULAR ARRAY CONE HYBRID [J].
PATTERSON, MS ;
FOSTER, FS .
ULTRASONIC IMAGING, 1983, 5 (03) :195-213
[12]   IMPULSE-RESPONSE AND PRESSURE NEARFIELD OF A CURVED ULTRASONIC RADIATOR [J].
PENTTINEN, A ;
LUUKKALA, M .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1976, 9 (10) :1547-1557
[13]   ERROR REDUCTION IN THROUGH TRANSMISSION TOMOGRAPHY USING LARGE RECEIVING ARRAYS WITH PHASE-INSENSITIVE SIGNAL-PROCESSING [J].
SCHMITT, RM ;
MEYER, CR ;
CARSON, PL ;
CHENEVERT, TL ;
BLAND, PH .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1984, 31 (04) :251-258
[14]   THE MALTESE CROSS PROCESSOR - SPECKLE REDUCTION FOR CIRCULAR TRANSDUCERS [J].
SMITH, SW ;
VONRAMM, OT .
ULTRASONIC IMAGING, 1988, 10 (03) :153-170
[15]   TRANSIENT RADIATION FROM PISTONS IN AN INFINITE PLANAR BAFFLE [J].
STEPANISHEN, PR .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1971, 49 (05) :1629-+
[16]   SPECKLE PATTERN CORRELATION WITH LATERAL APERTURE TRANSLATION - EXPERIMENTAL RESULTS AND IMPLICATIONS FOR SPATIAL COMPOUNDING [J].
TRAHEY, GE ;
SMITH, SW ;
VONRAMM, OT .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1986, 33 (03) :257-264
[17]  
TRAHEY GE, 1986, ULTRASONIC IMAGING, V8, P151, DOI 10.1016/0161-7346(86)90006-4
[18]  
von Ramm O. T., 1978, IEEE Transactions on Sonics and Ultrasonics, VSU-25, P340, DOI 10.1109/T-SU.1978.31040
[19]   STATISTICS OF SPECKLE IN ULTRASOUND B-SCANS [J].
WAGNER, RF ;
SMITH, SW ;
SANDRIK, JM ;
LOPEZ, H .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1983, 30 (03) :156-163