COMPARISON OF T-1 ESTIMATION TECHNIQUES IN CARDIAC MRI

被引:4
作者
WALKER, PM
MARIE, PY
DANCHIN, N
BERTRAND, A
机构
[1] Laboratoire de Biophysique, Faculté de Médecine, Université de Bourgogne
[2] Service de Cardiologie A, CHRU
[3] Service de Médecine Nucléaire, CHRU
关键词
CARDIAC MRI; T-1; TISSUE CHARACTERIZATION; QUALITY CONTROL;
D O I
10.1016/0730-725X(94)92351-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have shown that the use of a simple combination of inversion recovery/spin-echo (IR/SE) sequences provides undeniably superior precision in quantitative in vivo myocardium T-1 estimation than the standard multiple spin-echo approach. On a group of 25 healthy subjects, the T-1 dispersion was, respectively, 3.8% for the IR/SE combination and 19.6% for the best SE pair combination. Moreover, repeated measurements were carried out on seven of the volunteers in order to assess T-1 reproducibility. The mean intra-individual T-1 precision was found to be 2.8% for the IR/SE pair and 20.0% for the best SE pair. The in vivo imaging work was supported and corroborated by a thorough treatment of the theoretical T-1 errors. We also highlight the importance of quality control in quantitative MRI; a correction being required in the SE sequences to bring measured signal intensities into line with those predicted from simple theory.
引用
收藏
页码:43 / 50
页数:8
相关论文
共 14 条
  • [1] Aherne, Tscholakoff, Finkbeiner, Sechtem, Derugin, Yee, Higgins, MRI of cardiac transplants: The evaluation of rejection of cardiac allografts with and without immunosuppression, Circulation, 70, (1986)
  • [2] Wisenberg, Pflugfelder, Kostuk, McKenzie, Prato, Diagnostic applicability of MRI in assessing human cardiac allograft rejection, Am. J. Cardiol., 60, (1987)
  • [3] Prato, Drost, King, Keys, Wisenberg, Galland, Pflugfelder, Cardiac T<sub>1</sub> calculations from MR spin echo images, Magn. Reson. Med., 4, (1987)
  • [4] Prato, Drost, Keys, Laxon, Comissiong, Sestini, Optimisation of signal to noise in calculated T<sub>1</sub> images derived from two spin echo images, Magn. Reson. Med., 3, (1986)
  • [5] Pflugfelder, Wisenberg, Prato, Carroll, Turner, Early detection of canine myocardial infarction by MRI in vivo, Circulation, 71, (1985)
  • [6] van Dijkman, Doornbos, de Roos, van der Laarse, Postema, Matheijssen, Bruschke, van Voorthuisen, Cats, van den Wall, Improved detection of acute myocardial infarction by MRI using Gd-DPTA, Int. J. Cardiac Imaging, 5, (1989)
  • [7] Walker, Lerski, Mathur-de Vre, Binet, Yane, Preparation of agarose gels as reference substances for NMR relaxation time measurement, Magn. Reson. Imaging, 6, 2, (1988)
  • [8] Lerski, McRobbie, Straughan, Walker, de Certaines, Bernard, Multi-centre protocols and prototype test objects for the assessment of MRI equipment, Magn. Reson. Imaging, 6, 2, (1988)
  • [9] McRobbie, Lerski, Straughan, Orr, Slice profile effects and their calibration and correction in quantitative NMR imaging, Phys. Med. Biol., 32, 8, (1987)
  • [10] Walker, Marie, Mezeray, Besieres, Escanye, Danchin, Karcher, Mattei, Villemot, Bertrand, Synchronized inversion recovery-spin echo sequences for precise in vivo T1 measurement of human myocardium: A pilot study on 22 healthy subjects, Magn. Reson. Med., 29, pp. 637-641, (1993)