Extraction of a large vegetation and ICD lead using the Evolution Mechanical Dilator Sheath

被引:3
作者
Aksu, Tolga [1 ]
Durukan, Mine [1 ]
Guray, Umit [1 ]
Colak, Ayse [1 ]
机构
[1] Turkiye Yuksek Ihtisas Egitim & Arastirma Hastane, Kardiyol Klin, TR-06420 Ankara, Turkey
来源
TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY | 2011年 / 39卷 / 05期
关键词
Bacterial infections/complications; device removal/methods; electrodes; implanted; pacemaker; artificial;
D O I
10.5543/tkda.2011.01356
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 74-year-old male patient presented with complaints of fatigue, fever, and worsening health status. He had a 34-month history of cardioverter- defibrillator (ICD) implantation due to ischemic cardiomyopathy. Transthoracic and transesophageal echocardiography showed a mobile mass, 24x11 mm in size, consistent with a vegetation attached to the right ventricular ICD lead. Surgical removal of the electrode was considered to be highly risky for the patient, thus percutaneous removal was decided. Due to adhesions, manual traction of the lead with a standard stylet was ineffective. Complete lead extraction was accomplished using the Evolution Mechanical Dilator Sheath, which is a new mechanical sheath with a stainless steel bladed tip. No complications occurred during or after the procedure. Following a histopathologic diagnosis of vegetation, the patient received a six- week antibiotic therapy, after which a new ICD was implanted on the contralateral side. To our knowledge, this case represents the largest lead vegetation extracted by this new device. Although its safety should be validated by increasing number of cases, it seems that its simple use would make this device a new interesting tool among the instruments available for lead extraction.
引用
收藏
页码:403 / 406
页数:4
相关论文
共 7 条
[1]  
Bettig G., Doenecke P., Sen S., Complications with retained transvenous pacemaker electrodes, American Heart Journal, 98, 5, pp. 587-594, (1979)
[2]  
Wilkoff B.L., Love C.J., Byrd C.L., Bongiorni M.G., Carrillo R.G., Crossley III G.H., Et al., Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: This document was endorsed by the American Heart Association (AHA), Heart Rhythm, 6, pp. 1085-1104, (2009)
[3]  
Neuzil P., Taborsky M., Rezek Z., Vopalka R., Sediva L., Niederle P., Reddy V., Pacemaker and ICD lead extraction with electrosurgical dissection sheaths and standard transvenous extraction systems: Results of a randomized trial, Europace, 9, 2, pp. 98-104, (2007)
[4]  
Byrd C.L., Wilkoff B.L., Love C.J., Sellers T.D., Reiser C., Clinical study of the laser sheath for lead extraction: The total experience in the United States, PACE - Pacing and Clinical Electrophysiology, 25, 5, pp. 804-808, (2002)
[5]  
Kennergren C., Bucknall C.A., Butter C., Charles R., Fuhrer J., Grosfeld M., Tavernier R., Morgado T.B., Mortensen P., Paul V., Richter P., Schwartz T., Wellens F., Laser-assisted lead extraction: The European experience, Europace, 9, 8, pp. 651-656, (2007)
[6]  
Farooqi F.M., Talsania S., Hamid S., Rinaldi C.A., Extraction of cardiac rhythm devices: Indications, techniques and outcomes for the removal of pacemaker and defibrillator leads, Int J Clin Pract, 64, pp. 1140-1147, (2010)
[7]  
Bracke F., Complications and lead extraction in cardiac pacing and defibrillation, Neth Heart J, 16, SUPPL. 1, (2008)