PHASED-ARRAY DETECTORS AND AN AUTOMATED INTENSITY-CORRECTION ALGORITHM FOR HIGH-RESOLUTION MR-IMAGING OF THE HUMAN BRAIN

被引:103
作者
WALD, LL [1 ]
CARVAJAL, L [1 ]
MOYHER, SE [1 ]
NELSON, SJ [1 ]
GRANT, PE [1 ]
BARKOVICH, AJ [1 ]
VIGNERON, DB [1 ]
机构
[1] UNIV CALIF SAN FRANCISCO,GRAD GRP BIOENGN,SAN FRANCISCO,CA 94143
关键词
MAGNETIC RESONANCE IMAGING; PHASED ARRAY COILS; IMAGE INTENSITY CORRECTION; BRAIN;
D O I
10.1002/mrm.1910340321
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two- and four-coil phased array detectors were developed to increase the sensitivity and resolution of MR imaging of the human brain cortex, especially for detecting cortical dysplasias in pediatric epilepsy patients, An automated intensity correction algorithm based on an edge-completed, low-pass filtered image was used to correct the image intensity for the inhomogenous reception profile of the coils, Seven phased array coils were constructed and tested, The sensitivity of these coils was up to 600% higher at the surface of the cortex than that achieved with a conventional head coil and up to 30% greater at the center of the head. The sensitivity obtained was comparable with that of a conventional small surface coil, but extended over the larger dimensions of the array and previously inaccessible areas such as the top of the head, The advantages of the improved sensitivity are demonstrated with high resolution images of the brain.
引用
收藏
页码:433 / 439
页数:7
相关论文
共 29 条
  • [1] Hayes C.E., Hattes N., Roemer P.B., Volume imaging with MR phased arrays, Magn. Reson. Med., 18, pp. 309-319, (1991)
  • [2] Hayes C.E., Tsuruda J.S., Mathis C.M., Temporal lobes: surface MR coil phased‐array imaging, Radiology, 189, pp. 918-920, (1993)
  • [3] Roemer P.B., Edelstein W.A., Hayes C.E., Souza S.P., Mueller O.M., The NMR phased array, Magn. Reson. Med., 16, pp. 192-225, (1990)
  • [4] Hayes C.E., Roemer P.B., Noise correlations in data simultaneously acquired from multiple surface coil arrays, Magn. Reson. Med., 16, pp. 181-191, (1990)
  • [5] Hayes C.E., Dietz M.J., King B.F., Ehman R.L., Pelvic imaging with phased‐array coils: quantitative assessment of signal‐to‐noise ratio improvement, J. Magn. Reson. Imaging, 2, pp. 321-326, (1992)
  • [6] Foo T.K., MacFall J.R., Hayes C.E., Sostman H.D., Slayman B.E., Pulmonary vasculature: single breath‐hold MR imaging with phased‐array coils, Radiology, 183, pp. 473-477, (1992)
  • [7] Foo T.K., MacFall J.R., Sostman H.D., Hayes C.E., Single‐breath‐hold venous or arterial flow‐suppressed pulmonary vascular MR imaging with phased‐array coils, J. Magn. Reson. Imaging, 3, pp. 611-616, (1993)
  • [8] Hatabu H., Gefter W.B., Listerud J., Hoffman E.A., Axel L., McGowan J.D., Palevsky H.I., Hayes C.E., Konishi J., Kressel H.Y., Pulmonary MR angiography utilizing phased‐array surface coils, J. Comput. Assist. Tomogr., 16, pp. 410-417, (1992)
  • [9] Wald L., Vigneron D.B., Moyher S.E., Day M., Nelson S.J., (1994)
  • [10] Palmini A., Andermann F., Olivier A., Tampieri D., Robitaille Y., Focal neuronal migration disorders and intractable partial epilepsy: results of surgical treatment, Ann. Neurol., 30, pp. 750-757, (1991)