A COMPARISON OF ROUND-WINDOW AND TRANSTYMPANIC PROMONTORY ELECTRIC-STIMULATION IN COCHLEAR IMPLANT CANDIDATES

被引:25
作者
KILENY, PR
ZWOLAN, TA
ZIMMERMANPHILLIPS, S
KEMINK, JL
机构
[1] Department of Otolaryngology, University of Michigan Medical Center, Ann Arbor, MI
关键词
D O I
10.1097/00003446-199210000-00006
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
We compared within-subjects electrical thresholds and dynamic ranges obtained with direct round-window and transtympanic promontory stimulation carried out preoperatively in 12 patients who were candidates for a cochlear implant. Square waves with frequencies of 50, 100, 200, and 400 Hz were delivered in a 50% duty cycle to both sites in each patient. With the exception of threshold at 50 Hz (promontory thresholds were lower than round-window thresholds), there were no statistically significant differences for either thresholds or dynamic ranges between the two sites of stimulation. There was a general trend for round-window thresholds to be lower and dynamic ranges larger, especially for the higher frequencies of stimulation. Mean threshold slopes for the two sites of stimulation were nearly identical.
引用
收藏
页码:294 / 299
页数:6
相关论文
共 12 条
[1]  
BLACK FO, 1982, ANN OTO RHINOL LARYN, V91, P17
[2]  
BURIAN K, 1984, ACTA OTO-LARYNGOL, V41, P147
[3]  
CHOUARD CH, 1987, COCHLEAR IMPLANT CUR, P337
[4]   POST-STIMULATORY EFFECTS OF DIRECT-CURRENT STIMULATION OF THE COCHLEA ON AUDITORY-NERVE ACTIVITY [J].
COUSILLAS, H ;
PATUZZI, RB ;
JOHNSTONE, BM .
HEARING RESEARCH, 1988, 36 (01) :21-40
[5]   ELECTRICAL-STIMULATION OF THE ROUND WINDOW - A SELECTION PROCEDURE FOR SINGLE-CHANNEL COCHLEAR IMPLANTATION [J].
GRAY, RF ;
BAGULEY, DM .
CLINICAL OTOLARYNGOLOGY, 1990, 15 (01) :29-34
[6]  
JOHNSTONE BM, 1990, COCHLEAR IMPLANTS MO, P97
[7]   EFFECTS OF PREOPERATIVE ELECTRICAL STIMULABILITY AND HISTORICAL FACTORS ON PERFORMANCE WITH MULTICHANNEL COCHLEAR IMPLANT [J].
KILENY, PR ;
ZIMMERMANPHILLIPS, S ;
KEMINK, JL ;
SCHMALTZ, SP .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (07) :563-568
[8]  
LENARZ T, 1987, COCHLEAR IMPLANT CUR, P97
[9]  
MEYER B, 1984, ACTA OTO-LARYNGOL, P168
[10]  
SMOORENBURG GF, 1987, COCHLEAR IMPLANT CUR, P227