ACCURACY OF VELOCITY AND SHEAR RATE MEASUREMENTS USING PULSED DOPPLER ULTRASOUND - A COMPARISON OF SIGNAL ANALYSIS TECHNIQUES

被引:16
作者
MARKOU, CP
KU, DN
机构
[1] School of Mechanical Engineering, Georgia Institute of Technology, Atlanta
基金
美国国家科学基金会;
关键词
DOPPLER ULTRASOUND; ACCURACY; VELOCITY PROFILE; WALL SHEAR; FFT; HEMODYNAMICS; AUTOREGRESSIVE;
D O I
10.1016/0301-5629(91)90163-Q
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An experimental investigation was instituted to evaluate the performance of Doppler ultrasound signal processing techniques for measuring fluid velocity under well-defined flow conditions using a 10-MHz multigated pulsed ultrasound instrument. Conditions of fully developed flow in a rigid, circular tube were varied over a Reynolds number range between 500 and 8000. The velocity across the tube was determined using analog and digital zero crossing detectors and three digital spectrum estimators. Determination of the Doppler frequency from analog or digital zero crossing detectors gave accurate velocity values for laminar and moderately turbulent flow away from the wall (0.969 less-than-or-equal-to r less-than-or-equal-to 0.986). Three digital spectrum estimators, Fast Fourier Transform, Burg autoregressive method, and minimum variance method, were slightly more accurate than the zero crossing detector (0.984 less-than-or-equal-to r less-than-or-equal-to 0.994), especially at points close to the walls and with higher levels of turbulence. Steep velocity gradients and transit-time-effects from high velocities produced significantly larger errors in velocity measurement. Wall shear rate estimates were most precise when calculated using the position of the wall and two velocity points. The calculated wall shears were within 20%-30% of theoretically predicted values.
引用
收藏
页码:803 / 814
页数:12
相关论文
共 32 条
  • [1] Atkinson P, 1976, Ultrasound Med Biol, V2, P107, DOI 10.1016/0301-5629(76)90018-1
  • [2] BALASUBRAMANIAN K, 1979, THESIS SCH AEROSPACE
  • [3] BQKER DW, 1978, ULTRASOUND ITS APP 1, P161
  • [4] TOPOFLOW - DIGITAL FULL RANGE DOPPLER VELOCITY METER
    BRANDESTINI, M
    [J]. IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (05): : 287 - 293
  • [5] SOME CHARACTERISTICS OF LAMINAR-FLOW VELOCITY SPECTRA DETECTED BY A 20 MHZ PULSED ULTRASOUND DOPPLER
    BROWN, TD
    GABEL, RH
    PEDERSEN, DR
    BELL, LD
    BLAIR, WF
    [J]. JOURNAL OF BIOMECHANICS, 1985, 18 (12) : 927 - +
  • [6] BURG JP, 1975, THESIS STANFORD U ST
  • [7] DESPHANDE MD, 1977, THESIS SCH AEROSPACE
  • [8] MEASUREMENT OF FLUID TURBULENCE BASED ON PULSED ULTRASOUND TECHNIQUES .1. ANALYSIS
    GARBINI, JL
    FORSTER, FK
    JORGENSEN, JE
    [J]. JOURNAL OF FLUID MECHANICS, 1982, 118 (MAY) : 445 - 470
  • [9] TURBULENCE MEASUREMENTS WITH PULSED DOPPLER ULTRASOUND EMPLOYING A FREQUENCY TRACKING METHOD
    GIDDENS, DP
    KHALIFA, AMA
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (04) : 427 - 437
  • [10] GIDDENS DP, 1985, NONINVASIVE DIAGNOST, P58