DIFFERENTIAL EFFECT OF PANCURONIUM AT THE ADDUCTOR POLLICIS, THE 1ST DORSAL INTEROSSEOUS AND THE HYPOTHENAR MUSCLES - AN ELECTROMYOGRAPHIC AND MECHANOMYOGRAPHIC DOSE-RESPONSE STUDY

被引:15
作者
ENGBAEK, J [1 ]
ROED, J [1 ]
机构
[1] UNIV COPENHAGEN,GLOSTRUP HOSP,DEPT ANESTHESIOL & INTENS CARE,GLOSTRUP,DENMARK
关键词
MONITORING; NERVE; ELECTROMYOGRAPHY; MECHANOMYOGRAPHY; NEUROMUSCULAR BLOCKING AGENT; PANCURONIUM;
D O I
10.1111/j.1399-6576.1992.tb03540.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Cumulative dose-response curves were constructed from evoked compound electromyographic (EMG) recordings in man to compare the sensitivity to pancuronium of the adductor pollicis, the hypothenar and the first dorsal interosseous muscles. Also, the EMG and mechanomyography-based sensitivity of the adductor pollicis muscle were compared. The EMG and the mechanomyogram were evaluated in random sequence in each of 21 adult thiopental, fentanyl and diazepam anesthetized patients. The EMG-based ED50 were 36-38 mug . kg-1 with no differences between muscles. The EMG-based ED90 of the adductor pollicis and the hypothenar muscles were 62-65 mug . kg-1 compared to the 60 mug . kg-1 of the first dorsal interosseous muscle (P < 0.05). ED50 (34 mug . kg-1), and ED90 (56 mug . kg-1) obtained from the adductor pollicis mechanomyogram were significantly lower than those based on the EMG (P < 0.05). It is concluded that differences in sensitivity to pancuronium exist between the three muscles when evaluated from the EMG, and that the apparent sensitivity of a given muscle to a muscle relaxant may depend upon whether the response is evaluated using EMG or mechanomyography.
引用
收藏
页码:664 / 669
页数:6